60
Corona process | User Manual
Technical data
Corona process
Trigger IN
IN 0
GND
OUT 0
12 V
S1
Trigger OUT
Optical parameters
Spectrometer
High-power light source with
excellent optical properties and
internal referencing
Usable spectral range
380 – 1650 nm
Spectral resolution
(half tenth value width)
≤ 10 nm
Wavelength accuracy
≤ 1.0 nm
Wavelength drift
≤ 10 pm/K
Geometry
0° / 0°
Min. integration time
3 ms
Amplitude digitization
16 bits
Measuring distance
90 to 600 mm
Measuring spot
> 30 mm
Max. measurement frequency
1 Hz
Light source, lifetime
2 halogen lamps (1 redundant),
each 10,000 h
Warm-up time
< 30 min
Tab. 4 Optical parameters
Interface description
Tab. 5 Interface description