Care
61211-1310402
CONTURA
9-7
Inspection measures:
Object
Interval
Type of care
Stylus system
At least every week
Lint-free cloth; use a
cleaning agent if re-
quired
➤
Filter mat for ventilation of
the controller box
Quarterly
–
Clean by beating and
vacuuming
–
Replace the filter if re-
quired
➤
Filters for compressed air
As required
Check
➤
Care and cleaning
Measuring table
The contact surfaces of the workpiece or the clamping device and the
measuring table must be clean. Any particles beneath the contact sur-
face may lead to inaccurate measurements. Moreover, the surfaces of
the measuring table and of the workpiece may be damaged.
CAUTION
Risk of injury when a workpiece falls down due to an oily sub-
stance.
Crushing of feet and abrasions on legs.
ü
There must be
no lubricating film
on the measuring table and the
workpiece.
Remove any greasy or oily substances from the measuring table and
the workpiece.
1. Remove large particles with a vacuum cleaner.
2. Then clean the measuring table with a mild cleaning agent.
NOTE
Make sure that no cleaning agent residues remain on the measuring ta-
ble.
Summary of Contents for CONTURA
Page 55: ...4 61211 1310402 CONTURA 4 1 Technical data This chapter contains CMM 4 2 Technical data...
Page 62: ...CMM 4 8 61211 1310402 CONTURA...
Page 68: ...Installation 5 6 61211 1310402 CONTURA...
Page 90: ...Setting up the workpiece 6 22 61211 1310402 CONTURA 200 100 100 200 100 600 Size 9 12 Z...
Page 91: ...Setting up the workpiece 61211 1310402 CONTURA 6 23 600 100 600 100 600 100 Size 9 18 Z...
Page 92: ...Setting up the workpiece 6 24 61211 1310402 CONTURA 600 100 600 100 800 100 Size 12 18 Z...
Page 122: ...Terminating the measuring operation 7 26 61211 1310402 CONTURA...
Page 130: ...Service features 8 8 61211 1310402 CONTURA...
Page 150: ...2Glossary 2 61211 1310402 CONTURA...
Page 156: ...6Alphabetic index 6 61211 1310402 CONTURA...
Page 157: ......
Page 158: ...CONTURA Operating instructions 2022 04 29 61211 1310402...