0 Accessories and Optional System Expansions | 10.2 Aqua Stop
ZEISS
1
0.2 Aqua Stop
Purpose
The aqua stop protects the objectives and the nosepiece when working with liquid samples.
Position
The aqua stop is mounted on the nosepiece carrier.
1
0.2.1 Assembling the Aqua Stop
1
4
2
3
5
Fig. 8: Assembling the aqua stop
1
3
5
Cover cap
2
4
Cover disc
Lens hood
Collecting trough
Mounting holes
Hex Key, 3.0 mm
Parts and Tools
Prerequisite
ü
The stage is removed from the stand.
ü
All objectives and cover caps are removed from the nosepiece.
Procedure
5
1
2
3
4
. Remove the two cover screws from the mounting holes
.
3
. Place the collecting trough
on the nosepiece mount.
5
. Screw the collecting trough's two screws into the mounting holes
.
2
. Place the cover disk
the objectives.
at the nosepiece, matching the corresponding holes to those for
5
6
. Install the required objectives.
.
NOTICE
Make sure that each objective is enclosed up to the cover disk and that
no hood's upper edge forms a drip tray!
4
Pull a
small
lens hood
. Pull a
large
lens hood
over each objective of a front diameter of
16 to 22.5 mm
.
4
7
8
over each objective of a front diameter of
27.5 to 34 mm
.
1
. Close the unused nosepiece openings with cover caps
.
5
0