INSTALLATION INSTRUCTIONS AND FIRST-TIME SET-UP
ZEISS
Reflector turret
Axio Observer
54
431004-7244-001
12/2016
4.11
Reflector turret
4.11.1
Fitting the reflector turret
The reflector turret (manual or motorized) is
inserted into the stand from the right.
Switch the instrument off before
inserting the motorized reflector turret.
Close the RL shutter when changing
the reflector turret to prevent the
emission of refracted light.
•
Rotate the locking lever (Fig. 34/
2
) downwards
and open the cover (Fig. 34/
1
).
•
Insert the loaded or unloaded reflector turret
(Fig. 34/
3
) into the opening beneath the
nosepiece as far as it will go.
•
Close the cover flap (Fig. 34/
1
) and rotate the
locking lever (Fig. 34/
2
) upwards.
Instead of the reflector turret the dual filter wheel mot. can be inserted into the stand. See
Quick Reference Guide “Filter wheel excitation 8-pos. mot. for filters d = 25 mm; CAN and Dual
filter wheel mot. for beam splitting and emission; CAN” (452358-7044-001).
Fig. 34
Fitting the reflector turret