START-UP
Axio Imager
Attaching the condenser carrier to …
Carl Zeiss
M70-2-0020 e 06/2009
430000-7344-001
43
3.16
Attaching the condenser carrier to
the transmitted-light/reflected-
light stage carrier
•
2
) to adjust the guide
of the condenser carrier (3-18/
3
) in such a way
4
) are accessible.
•
Attach the condenser carrier to the stage carrier
(3-18/
1
). Push it to the upper stop in a firm and
straight manner, and tighten the two screws.
3.17
Changing the fine-drive knob or
the fine-drive disk on the stand
The fine-focusing drive to the right and left on the
stand can be provided either with a fine-drive knob
or a fine-drive disk.
To change the fine-drive knob or the fine-drive
disk, proceed as follows:
•
Turn the coarse-drive knob on the relevant side
of the stand in such a way that the mounting
hole (3-19/
2
) points upward.
•
Insert the ball-headed screwdriver SW 1.5
(3-19/
1
) into the mounting hole and loosen the
locking screw of the fine-drive knob or the fine-
drive disk.
•
Pull the fine-drive knob or the fine-drive disk
sideways off the shaft.
•
Slide the fine-drive disk or the fine-drive knob
onto the shaft, so that the recess (3-19/
3
or
4
)
is located under the mounting hole.
•
Tighten the locking screw.
Fig. 3-18
Attaching the condenser carrier to
the stage carrier
Fig. 3-19
Changing the fine-drive knob or
fine-drive disk