18
Changing the motor
AURA
®
handheld NIR - Rev. 2
Maintenance and Repair Manual | AURA
®
handheld NIR - Rev. 2 | EN_40_070_0030I
6 Changing the motor
•
The display has been removed (see section “3 Changing the display”,
steps 1 to 6).
•
The lamp holder with halogen lamp has been removed
(see section “4 Replacing the halogen lamp”, step 1).
•
Put the following spare parts ready:
– Motor with cable, order no. 000000-2326-676
•
Put the following tools and equipment ready:
– TX 8 Torx screwdriver
–
2 mm flat tip screwdriver
1
Remove the fastening screws of the front panel
Lay down the device with its back
1
facing upwards.
Unscrew the seven Torx screws
2
(3x on rear, 2x on top, 2x on bottom)
from the housing. Use a TX 8 Torx screwdriver for this.
Fig. 12
2
1
Remove the fastening screws on the front panel
Hold the front panel and turn the device so that you can lay it back down
with the open device side facing upwards.
Requirements
Procedure