Control the position of the lens
within the channel of the closed
cartridge. Haptic must be freely
movable.
Place the lens in the center of the
opened cartridge.
AT.Shooter A2-2000
Preparing Monofocal and Monofocal Toric MICS IOLs for use with the reusable
AT.Shooter A2-2000 injector
Carl Zeiss Meditec AG
Goeschwitzer Str. 51–52
07745 Jena
Germany
www.zeiss.com/iol
www.zeiss.com/med/contacts
1
4
7
2
5
8
3
6
9
Please refer to our web pages for the most up-to-date references.
ACM2 Cartridge Set
AT.Shooter A2-2000
*Carl Zeiss Meditec recommends the use of ZEISS OVDs.
Place a small amount of OVD* in the
cartridge tip. Place a line of OVD in
each channel of the cartridge.
Do not touch the front tip of the
cartridge.
Pull the lens gently from its holder
using a forceps with fine tips.
Do not let the IOL dry up.
Do not use the fluid from the bottle
for ocular irrigation.
Maintain the IOL in place with the
folding forceps while closing the
cartridge. Close the cartridge until
you hear a click.
Mount the tip holder onto shooter
tip. Insert the plunger into the recess
of the blue silicone tip and remove
the tip from its holder by retracting
the plunger.
Place the cartridge into the loading
deck of the shooter and push it
forward.
Gently push the plunger with the
blue silicone tip mounted on it
straight into the cartridge channel
and start to push the lens forward.
Check the correct positioning of the
lens. Gently push forward the lens
until it becomes visible in the tip of the
cartridge. Start implantation of lens.
AT.Shooter A2-2000
0297
ACM2 Cartridge Set