AT.Shooter A1-2000 with  
VISCOJECT-BIO 2.2 Cartridge Set 

Preparing the compatible IOLs for use.

Carl Zeiss Meditec AG

Goeschwitzer Str. 51–52
07745 Jena
Germany 
www.zeiss.com/iol
www.zeiss.com/med/contacts

AT.Shooter A1-2000

Place a small amount of OVD* in the 
cartridge tip. Place a line of OVD in 
each channel of the cartridge. 
Do not touch the front tip of the 
cartridge.

Pull the lens gently from its holder 
using a forceps with fine tips. 
Do not let the IOL dry up. 
Do not use the fluid from the bottle 
for ocular irrigation.

Place the lens in the center of the 
opened cartridge and move it a 
little forward towards the channel 
opening.

1

2

3

Maintain the IOL in place with the 
folding forceps while closing the 
cartridge. Close the cartridge until 
you hear a click.

Close the cartridge without clamping 
the lens. Control the position of the 
lens within the channel of the closed 
cartridge. Haptic must be freely 
movable.

Mount the tip holder onto shooter 
tip. Insert the plunger into the recess 
of the blue silicone tip and remove 
the tip from its holder by retracting 
the plunger.

4

5

6

Place the cartridge into the loading 
deck of the shooter and push it 
forward.

Gently push the plunger with the 
blue silicone tip mounted on it 
straight into the cartridge channel 
and start to push the lens forward.

Check the correct positioning of the 
lens. Gently push forward the lens 
until it becomes visible in the tip of the 
cartridge. Start implantation of lens.

7

8

9

0297

CT 47LC 
CT 47S 
CT 37A 
AT LISA 801

Please refer to our web site for the most up-to-date references. 

*Carl Zeiss Meditec recommends the use of ZEISS OVDs.

VISCOJECT

-BIO 2.2 Cartridge Set

AT.Shooter A1-2000