33
Calibrating the compensation data
AURA
®
handheld NIR - Rev. 2
EN_40_070_0030I | AURA
®
handheld NIR - Rev. 2 | Maintenance and Repair Manual
11
Photometric verification – Gray reference measurement
Place the gray reference on the centering aid.
Make sure that the reference is correctly aligned with the attachment
(the nose of the reference should slot into the groove of the attachment).
Click
OK
to start the measurement.
12
Photometric verification – White reference measurement
Place the white reference on the centering aid.
Click
OK
to start the measurement.