Calibrating the compensation data
AURA
®
handheld NIR - Rev. 2
Maintenance and Repair Manual | AURA
®
handheld NIR - Rev. 2 | EN_40_070_0030I
9
Referencing – Calibration of cavity (black reference)
Place the black reference on the centering aid. Make sure that the markings
on the reference and the centering aid are aligned.
Click
OK
to start the calibration.
10
Photometric verification – Cavity measurement (black reference)
Place the black reference on the centering aid.
Click
OK
to start the measurement.
Click
to repeat the measurement.
Click
to proceed to the next step.
= Repetition of last
part of program