0004-000067
49
Process Variable
Type
Values
Description
setStatusOperEnable
getStatusOperEnable
mbboDirect
mbbiDirect
0000
16
to
FFFF
16
Operation register bit enable mask
Bit 0 = calibrating
Bit 1 = settling
Bit 2 = ranging
Bit 3 = sweeping
Bit 4 = measuring
Bit 5 = waiting for trigger
Bit 6 = waiting for arm
Bit 7 = unused
Bit 8 = trigger event
Bit 9 = data capture event
Bit 10 = limit test event
Bit 11 = auto download event
Bits 12-15 = unused
StatusOperCond mbbiDirect
0000
16
to
FFFF
16
Operation condition register:
Bit 0 = calibrating
Bit 1 = settling
Bit 2 = ranging
Bit 3 = sweeping
Bit 4 = measuring
Bit 5 = waiting for trigger
Bit 6 = waiting for arm
Bit 7 = unused
Bit 8 = trigger event
Bit 9 = data capture event
Bit 10 = limit test event
Bit 11 = auto download event
Bits 12-15 = unused
StatusPreset
bo
0 or 1
Sets the status reporting event enable data
structures to a known state. The condition
and event register contents are not
affected. All device-dependent status
registers which cascade events into the
Questionable Status and the Operation
Status Registers are enabled by setting
those device-dependent event enable
registers to 7FFF
16
(the 15 LSBs set). The
IEEE-488.2 mandatory status data
structures are disabled by setting the
Questionable Status and Operation Status
event enable registers to 0000
16
. The
Status Byte and Standard Event Status
Registers as defined by IEEE 488.2 are not
affected