10

Product Properties

1.Operation of the product is simple and convenient.

2.The product is small in size, light in weight and portable.

3.The product can operate continuously for 17 hours with two brand new AAA

batteries.(The operation time may vary due to the different performance of batteries)

4.Low voltage prompt will appear on the display when the battery voltage is lower

than the minimum value of normal working voltage range.

5.The product will automatically shut down when there is no signal detected for

about eight seconds.

Product Operation Scope

The YUWELL® Finger Pulse Oximeter is designed for fingers(not thumb)

between 0.3 and 0.9 inch (0.8-2.3cm) thick. And the finger shall be inserted into the

sensor position which is in the middle of the device.

The Finger Pulse Oximeter is NOT design for newborns and infants.

Product Operation Steps

1.Install two AAA batteries into battery cassette before closing the cover.

2.Nip the clamp as diagram.(See Figure)

3.Plug one finger into rubber hole of the Oximeter

(it is best to plug the finger thoroughly) before

releasing the clamp.

4.Press the switch button one time on the front panel.

5.Do not tremble while the oximeter is working.

It's better that the whole body be in still status.

Note: For normal use after long interruptions, refer to

the product operation steps.

6.Read corresponding data from display screen.

Battery Installation

1.Pull the battery cover horizontally.

2.Put two AAA batteries into battery cassette in

right polarities and push the cover back.

Notes: Battery polarities must be correct. Otherwise,

Summary of Contents for YX310

Page 1: ...1 YUWELL Finger Pulse Oximeter YX310 Please read the user manual carefully before using The picture is for reference only please refer to the actual product...

Page 2: ...ntify interference and the elimination of the following measures 1 Shut down the equipment in the vicinity and then re open in order to identify interference equipment 2 To change the direction or loc...

Page 3: ...dex finger middle finger and ring finger instead of small finger 22 Warning Do not use the oximeter if it shows signs of damage or is suspected of being damaged Inaccurate or no readings may result fr...

Page 4: ...wer levels 34 This product can be operated by the patient or by others to measure the patient s PR and SpO2 The maintenance operation and maintenance methods are the same This product contains batteri...

Page 5: ...yhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology so that two beams of different wavelength of light red light and infrared light can be...

Page 6: ...dication for Use Intended use The YUWELL Finger Pulse Oximeter is a non invasive non sterile reusable spot checking device which can measure and display SpO2 and pulse rate through finger It is intend...

Page 7: ...approximately 2 thirds of the values were at over or below the accuracy Arms value Pulse rate 25bpm 250bpm accuracy Arms 1 or 1bpm larger 5 Anti interference ability of ambient light Deviation in blo...

Page 8: ...P22 17 Description of oximeter application management YX310 is equipped with Bluetooth function Bluetooth communication protocol module enables the oximeter to be equipped with Bluetooth connection an...

Page 9: ...The Skin tones included in the study were as follows 3 subjects with dark pigmentation 5 subject with very light pigmentation The remaining subjects with light medium skin tones of Asian origins 5 Gra...

Page 10: ...all be inserted into the sensor position which is in the middle of the device The Finger Pulse Oximeter is NOT design for newborns and infants Product Operation Steps 1 Install two AAA batteries into...

Page 11: ...the hanging hole 2 Thread thicker end of the lanyard through the threaded end before pulling it tightly 3 Install as the figures show See Figure Maintenance and Storage Under normal conditions there i...

Page 12: ...r soft cloth moistened with water 3 After cleaning wipe off the water with a soft cloth 4 Allow the oximeter to air dry The most commonly used hospital cleaning and non corrosive liquid detergent can...

Page 13: ...oblem exiting in the product please go to a hospital timely for exact diagnosis 3 Nail polish or paste manicure 3 Remove the nail polish or discharge manicure when measuring SpO2 or PR is shown unstab...

Page 14: ...on The following degradations if associated with BASIC SAFETY or ESSENTIAL PERFORMANCE shall not be allowed component failures changes in programmable parameters or settings reset to default settings...

Page 15: ...ME SYSTEMS Table 4 For ME EQUIPMENT and ME SYSTEMS that are not LIFE SUPPORTING RS Guidance and manufacturer s declaration Radiated Emission Emissions test IEC60601 test level Compliance level Result...

Page 16: ...Band 13 17 Pulse modulationb 217Hz 0 2 0 3 9 745 780 810 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 Pulse modulationb 18Hz 2 0 3 28 870 930 1720 1700 19 00 GSM 1800 CDMA 1900 GSM 1900...

Page 17: ...ent The device has been evaluated to meet general RF exposure requirement The device can be used in portable exposure condition without restriction Note This equipment has been tested and found to com...

Page 18: ...Card Thank you very much for using our products Product name Finger pulse oximeter Model YX310 S N MFG DATE JIANGSU YUYUE MEDICAL EQUIPMENT SUPPLY CO LTD Manufacturer Address No 1 Baisheng Road Develo...

Reviews: