![Yuwell YE680B User Manual Download Page 21](http://html1.mh-extra.com/html/yuwell/ye680b/ye680b_user-manual_3753448021.webp)
Technical parameters
Technical parameters
38
39
YE680B
LCD digital display
Oscillation mensuration
0-300mmHg
0-450mmHg
40-200 time/min
Pressure:±3mmHg(±0.4kPa)
Pulse:±5% of reading value
=Class II (optional AC adapter)
=Type BF application part
4X1.5v AA batteries
or 6V/ 600mA AC adapter
300 times
22~45cm
IP 21
Approx.330g
Approx.127x93x74 (mm)
+5℃ to +40℃/15% to
90%
RH (non-condensing)
-20℃ to +55℃/15% to 90%
RH (non-condensing)
5 years or 50 thousand times
Model
Display
Measuring
method
Measuring range
Cuff Pressure
Pulse rate
Precision
Electric
classification
Power supply
Battery life
Cuff size
IP Classification
Service life
Weight
Dimension
Operating
temperature
/Humidity
Storage temper-
ature /Humidity
Operating/Storage
atmospheric
pressure
70kPa to 106kPa
There is the potentia risk of radio frequency
interference between the device and other
devices. If there is, please find out the
problems and take the following measures:
(1) Turn off the device, and turn on again.
(2) Change the direction of the device.
(3)Keep
the product
away from the
interferential devices.
Part
Material
Rear Cover
Top Cover
ABS
PC
Magic paster
edge cloth
Cuff
Air tube
Air plug connect
PVC
ABS
Nylon
Polyester cotton
The contact materials detail of product
Statement!
The
SPHYGMOMANOMETER
was
clinically
investigated
according
to
the
requirement
of
ISO
81060
-
2
:
2013
.
The SPHYGMOMANOMETER complies with
IEC 80601-2-30.