Caution: 

During your therapy, you can go to the clinical menu by Pressing and holding START/STOP button and 
dial at the same time for three seconds. Section 11.2 clinical menu tells the method to change all the 
parameters.

12. Caring for your device

It is important that you regularly clean your device to make sure you receive optimal therapy. The 
following sections will help you with disassembling, cleaning, checking, reassembling your device. 

 

12.1 Disassembling

1. Pull the power connector out of device

2. Hold the cuff of the air tubing and gently 

pull it away from device

  

3. Press the Slide switch, and take out the water 

tank.

4. Pour out the left water in the water tank.

6

5

5. Press the water tank gently and pull it away 

from the device.

6. 

swivel of the mask, then gently pull apart.

 

Notice: Don't pull the threaded portion 

         of the air tubing

   

 

12.2 Cleaning

WARNINGS!

•  Regular cleaning of the device and its accessories is very important for the prevention of respiratory 

infections.

•  To avoid electric shock, always unplug the device before cleaning.
•  Use mild soap that is nontoxic to humans.
•  Before cleaning, check whether the device has been disconnected from the power supply, whether 

the power cord has been unplugged, and whether the water tank of the device has cooled down. Make 
sure the plate has cooled down to room temperature, so you do not get burned.

•  Do not open or modify the device. There are no user serviceable parts inside. Repairs and servicing 

should only be performed by an authorized service agent.

CAUTIONS!

•  Overheating of the materials could lead to early fatigue of these materials.
•  Do not clean the device with bleach, chlorine, ozone, UV light, aromatic solution, moisturizing factor, 

antibacterial soaps, sesame oil, or any other methods outside of this manual.

• 

Summary of Contents for BreathCare PAP

Page 1: ...zhou Science Technology Town 215163 Suzhou Jiangsu PRC TEL 86 0512 67373001 Web address www yuyue com cn Shanghai International Holding Corp GmbH Europe IFU V 02 version 04 Prepared date 2022 06 Breat...

Page 2: ...f a physician 1 Indications for use 01 2 Contraindications 01 3 Warnings 01 4 Precautions 03 5 Adverse reactions 04 6 Package table 05 7 Picture and explanation for product 05 8 Explanation of button...

Page 3: ...the device where it can be bumped or where someone is likely to trip over the power cord Do not block the air tube and or air inlet of the device while in operation otherwise it could cause overheatin...

Page 4: ...age the device Use only vented masks recommended by Yuwell or by your doctor with this device Fitting the mask without the device blowing air can result in rebreathing of exhaled air Make sure that th...

Page 5: ...tion of button START STOP button Press to start stop therapy Dial Turn to navigate the menu and press to select an option Turn to adjust the options and press to save your choice 9 Installation The fo...

Page 6: ...utomatically after two minutes of no action You can press any button to turn it back on Caution When the level of the humidity is 1 6 and the device is delivering gas pressure the heating plate will h...

Page 7: ...6 55 This temperature has been tested with empty water tank with 10 minutes preheating time To adjust the humidity level Highlight the Humidity press dial to enter the humidity page Press the dial and...

Page 8: ...se or decrease therapy pressure each step is 0 5 cmH2O hPa After setting therapy pressure press dial to save your change Highlight the Back press dial to back to home page You can set Max Pressure Min...

Page 9: ...t Notice Don t pull the threaded portion of the air tubing 12 2 Cleaning WARNINGS Regular cleaning of the device and its accessories is very important for the prevention of respiratory infections To a...

Page 10: ...daily cleaning of the water chamber 12 2 2 Cleaning the Enclosure Wipe the surface of the device with a soft slightly damp cloth CAUTIONS The device can only be used after the enclosure is dry so tha...

Page 11: ...f the mask is not suitable See your mask user guide to check Please refer to the nose measurement card to choose the appropriate mask size I am getting a dry or blocked nose Humidity level may set too...

Page 12: ...contact your provider or Yuwell ERROR 8 There is a mistake of blower Please contact your provider or Yuwell ERROR 9 There is a mistake of RTC loss power Please adjust the time of the device or contac...

Page 13: ...leakage 2 L min at the maximum rated pressure The performance of the device at set pressure according to ISO 80601 2 70 2015 is shown below Test pressures hPa 4 8 12 16 20 CONNECTION PORT L min 100 1...

Page 14: ...d up to 15 degrees 17 2 Disposal statement Warning Please contact local authorities your provider or Yuwell to determine the proper method of disposal of this device 18 Limited warranty Yuwell warrant...

Page 15: ...s Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 0 UT 0 5 cycle At 0 45 90 135 180 225 270 and 315 0 UT 1 cycle 70 UT 25 30 cycle At 0 0 UT 250 300...

Page 16: ...PMENT or ME SYSTEM may be reduced to 1 m The 1 m test distance is permitted by IEC 61000 4 3 a For some services only the uplink frequencies are included b The carrier shall be modulated using a 50 du...

Page 17: ...gn Date Thislimitedwarrantydoesnotcover Any damage caused as a result of improper use abuse modi cation or alteration of the product Repairs carried out by any service organizations that have not been...

Page 18: ......

Reviews: