background image

14

15

OPERATING & INSTALLATION

. TURNNING OFF

Take  off  nasal  oxygen  cannula  from  oxygen  outlet  first, 
close power switch, and then cut off power source.

Symbol  

Description

Alternating current

Symbol  

Description

Consult the manual

ClassⅡEquipment

Type BF Applied part

ON(power, connection 
to the mains)

OFF(power, disconnection 
from the mains)

KEEP UP

FRAGILE

KEEP DRY

. SYMBOLS

OPERATING & INSTALLATION

The alarm signal

1.When  oxygen  concentration

  ≥

82%  the  green  indicator 

lights, LCD shows " elapsed time

 -–

 " the machine is in good 

condition.

2.When  oxygen  concentrations

 

82%  the  yellow  indicator 

light,  LCD  shows  "elapsed  time 

-–

  ".  Please  contact  the 

supplier  immediately.  Users  can  continue  to  use,  and  please 
ensure that backup oxygen is nearby.

NOTE:

 The minimum operating time is 30 minutes.

3.Pressure,  cycle  failure  alarm--the  red  indicator  lights  with 
continuous alarm sound, LCD shows "E1 or E2" machine stops 
working.  Please  shut  down  immediately  and  use  backup 
oxygen, and contact the supplier immediately.

4.Compressor  failure  alarm--the  red  indicator  lights  with 
continuous  alarm  sound,  LCD  shows  "E3  or  E4",  machine 
stops  working.  Please  shut  down  immediately  and  use 
backup oxygen, and contact the supplier immediately.

5

.In case of power loss alarm, the red indicator will be turned 

on, an alarm will be sent, no display, and the machine must 
be shutdown. Please check the power input. 

4

IPX0

No open flam

e

Unprotected

Stacking imitation 

Summary of Contents for 7F-5A

Page 1: ...ely before using JIANGSU YUYUE MEDICAL EQUIPMENT SUPPLY CO LTD Sales Center Address Huanyuan East Road No 1 Xuzhuang Software Park Nanjing Jiangsu Province P R China 210000 Manufacturing Address YunYa...

Page 2: ...e concentrator should always be kept in the upright to prevent cabinet damage when transporting Only the techncian can open the oxygen concentrator s cabinet BEFORE INSTALLATION STORAGE PLACE 1 You ma...

Page 3: ...ly ignited and burn with great intensity in oxygen enriched air Failure to observe this warning can result in severe fire property damage and cause physical injury or DEATH For optimum performance do...

Page 4: ...ctive pneumonia disease rehabilitation and hypoxia conditions Oxygen could improve body function especially suit for elderly pregnant women physique students etc who got physiological hypoxia It may a...

Page 5: ...work continuously 16 Minimum operating time 30minutes 20 NORMAL OPERATION CONDITIONS Temperature range 10 40 50 104 19 Enclosures Protection Classification IPX0 10 PowerSupply AC220V 22V 50Hz 1Hz 14 A...

Page 6: ...rosive gas and ina well ventilated room avoid jolt and inverted recumbent transport FEATURES NOTE Unless the oxygen concentrator is to be used immediately otherwise it should be stored in the containe...

Page 7: ...counter clockwise Fill the bottle with pure water or distilled water to the level indicated by the manufacturer Oxygen outlet Oxygen inlet Safety valve on the inside Cover Tube OXYGEN ABSORBING OPERA...

Page 8: ...pull out the nebulizing cover on the nebulizing interface Figure 7 Figure 7 OPERATING INSTALLATION NOTE the time for nebulizing should follow the doctor s advices When nebulizing completing please tu...

Page 9: ...ime Please contact the supplier immediately Users can continue to use and please ensure that backup oxygen is nearby NOTE The minimum operating time is 30 minutes 3 Pressure cycle failure alarm the re...

Page 10: ...7F 5AW Pull out the air hose pull the mouthpiece downward pull the nebulizing cup s cap empty remained medicinal liquid in the nebulizing cup then wash the air hose mouthpiece nebulizing cup s cap neb...

Page 11: ...the power cord plug into the outlet firmly 2 There is no power output from the outlet 2 Move it to the outlet with power output 3 There is insufficient power from the outlet 3 Do not use the extension...

Page 12: ...ds and E3 is shown on the display screen 1 The compressor circuit is open circuit 1 Shutdown the machine and contact the supplier immediately Solution Probable cause Symptom In atomization operation t...

Page 13: ...al oxygen cannula Filter Bacilli filter OTHER ATTENTION ITEMS OTHER ATTENTION ITEMS Low power supply Transformer Fuse Electronic control circuit Assembled valve Assembled valve Compressor Fan ELECTRIC...