background image

Floor should be wood, 

concrete, or ceramic 

tile. If floors are 

covered with synthetic 

material, the relative 

humidity should be at 

least 30% RH.

Mains power quality 

should be that of a 

typical commercial

and/or hospital 

environment.

Mains power quality 

should be that of a 

typical commercial

and/or hospital 

environment.

±6 kV contact

±8 kV air

±2 kV for 

power

supply lines

Not applicable

Surge

IEC 

61000-4-5

±1 kV 

line to line 

±2 kV 

line to earth

±1 kV 

line to line 

not applicable

<5 % 
(>95 % dip in 

)

for 0.5 cycle

40 % U

T

(60 % dip in U )

T

for 5 cycles

70 % U

T

(30 % dip in U )

T

for 25 cycles

<5 % U

T

(95 % dip in U )

T

for 5 sec.

U

T

U

T

<5 % 
(>95 % dip in 

U )

T

for 0.5 cycle

40 % U

T

(60 % dip in U )

T

for 5 cycles

70 % U

T

(30 % dip in U )

T

for 25 cycles

<5 % U

T

(95 % dip in U )

T

for 5 sec.

U

T

Voltage dips,

short 

interruptions

and voltage

variations on

power supply

IEC 

61000-4-11

±6 kV contact

±8 kV air

±2 kV for 

power

supply lines

±1 kV for

input/output 

lines

Electrostatic

discharge

IEC 

61000-4-2

Electrical fast

transient/burst

IEC 

61000-4-4

Mains power quality 

should be that of a 

typical commercial

and/or hospital 

environment.

If the user of the 

aspirators 7E-B5 

requires continued

operation during power 

mains interruption, it is

recommended that the 

7E-B5 be powered 

from an uninterruptible 

power supply

or battery.

Compliance

level

Electromagnetic

environment

 guidance

IEC 60601

Test level

Immunity test

Table 2 Guidance and manufacturer's declaration 

 electromagnetic immunity

7E-B5 is intended for use in the electromagnetic environment specified 

below. The customer or the user should assure that it is used in such 

environment

Guidance and manufacturer's declaration 

 electromagnetic immunity

3 A/m

Power frequency 

magnetic fields should 

be at levels 

characteristic of a 

typical location in a 

typical commercial or 

hospital environment

Power 

frequency

(50/ 60 Hz)

magnetic field

IEC 

61000-4-8

3 A/m

3 A/m

Note:   is the A.C. mains voltage prior to application of the test level.

U

T

Table 3 Guidance and manufacturer's declaration 

 electromagnetic immunity

Compliance

level

Electromagnetic

environment

 guidance

IEC 60601

Test level

Conducted 

RF

IEC

61000-4-6

Radiated 

RF

IEC 

61000-4-3

3 V rms

150 kHz ~

80 MHz

3 V/m

80 MHZ to

2.5 GHz

3 V rms

3 V/m

Immunity

test

7E-B5  is intended for use in the electromagnetic environment specified 

below. The customer or the user of these YUWEII aspirators should 

assure that it is used in such environment.

Guidance and manufacturer's declaration 

 electromagnetic immunity

Portable and mobile RF commu-

nications equipment should be 

used no closer to any part of the 

YUWEII aspirators 7E-B5 

including cables, than the 

recommended separation 

distance calculated from the 

equation appropriate to the 

frequency of the transmitter.

Recommend separation distance

d=1.2

d=1.2       80MHz to 800 MHz

d=2.3       800 MHz to 2.5 Ghz

where P is the maximum output 

powerrating of the transmitter in 

watts (W) according to the 

transmitter manufacturer and d is 

the recommended separation 

distance in meters (m).

Field strengths from fixed RF

transmitters as determined by an

electromagnetic site survey, a 

should be less than the compli-

ance level in each

frequency range. b Interference 

mayoccur in the vicinity of 

equipment markedwith he 

following symbol: 

13

12

Summary of Contents for 7E-B5

Page 1: ...Phlegm Suction Unit JIANGSU YUYUE MEDICAL EQUIPMENT SUPPLY CO LTD Sales Center Address Huanyuan East Road No 1 Xuzhuang Software Park Nanjing Jiangsu Province P R China 210000 Manufacturing Address Yu...

Page 2: ...only describes the usage of the product Do not use accessories other than the manufacturer s recommendation This will reduce the performance of the machine 4 Please do not use the machine and return...

Page 3: ...agram 2 Product Features Installing and Commissioning The customer shall carefully inspect if the appearance of product is good and the varieties quantities of the attachments are in conformity with t...

Page 4: ...he holder will not ascend when the suction tube conductor is placed again into the liquid as suctioned and for Item 2 the liquid level still ascends Thus it is required to observe carefully and lift i...

Page 5: ...in the air filter which leads to gradually darkening of the color of filter diaphragm and obviously reducing or even disappearing of suction force at the inlet of tube while the negative pressure indi...

Page 6: ...ng and application Wipe the case outer surface with lightly wet rag already soaked in the disinfectant and prevent any liquid seeping into the pump Never wipe the places marked with letters and patter...

Page 7: ...Operator due to electromagnetic disturbances Attachments 11 10 1 Suction conductor length 2m 7 12 1 pc 2 Suction catheter F8 F12 1 pc respectively of child adult 3 Power cord 1 pc 4 User s manual 1 pc...

Page 8: ...customer or the user should assure that it is used in such environment Guidance and manufacturer s declaration electromagnetic immunity 3 A m Power frequency magnetic fields should be at levels charac...

Page 9: ...etic environment in which radiated RF disturbances are controlled The customers or the users of these YUWELL aspirators can help prevent electromagnetic interference by maintaining a minimum distance...

Reviews: