![YUSHI YUT-2800 Manual Download Page 24](http://html1.mh-extra.com/html/yushi/yut-2800/yut-2800_manual_3745659024.webp)
YUSHI INSTRUMENTS
21
3.7.5 Communication
The flaw detector can exchange data with PC. (For more information, see Chapter 8). The default status
is on.
3.8 Operation of Other Special Functional Keys
3.8.1 Automatic Gain
Move the gate to lock the echo that you want, as shown in Figure 3-1. Press to start the
automatic gain. The instrument will repeatedly compute the echo inside the gate automatically and adjust
the gain automatically until the amplitude height of the largest echo is set to 80% of the screen height, as
shown in Figure 3-2.
Note: the wave height after the automatic gain is related to the automatic wave height setting in the
setting menu. If the automatic wave height is set as a certain value (in between 20%-80%), the wave
height after the automatic gain is the height of the value on the screen.
Note: When the DAC curve has been made, press the automatic gain, the instrument will
automatically adjust the compensating gain so as to make the echo inside the gate reach 80% of
the screen. The changes of echoes do not mean the changes of DAC curve. However, if the option is
on the parameter menu and on the gain option, press the automatic gain, the instrument will
adjust the gain automatically. The echo and the DAC cure change simultaneously.
Figure 3-1
Figure 3-2