3. Functions and Applications 

3.1 Technical Specifications 

Measuring Method: Leeb Hardness Testing method 

Hardness Scale: HL, HB, HRB,HRC,HV,HS, σ

b.

 

Measuring  Range  :  HLD  (200-960)  HRC

19.8-68.5

  HB

30-651

  HV

80-976

  HS

26.4-99.5

  HRB

13.5-100

σ

b

 (375-2639). 

Impact device: D type impact device (Standard); Optional C/G/DC/DL/D+15 

Accuracy: 

±6HLD (HLD=800), Repeatability 6HLD (HLD=800) 

Measuring direction: 360 degree by manual setting 

Material  :  Steel&  Cast  Steel,  Stainless  Steel,  GC  IRON,  NC  IRON,  Cast 

Aluminum alloy, Macht metal, Copper-tin alloy, Brass 

Resolution: 1HL, 1HV, 1HB, 0.1HRB, 0.1HRC, 0.1HS 

Display: LED with backlight. 

Memory: 100 groups (each group include 1-7 testing result and 1 AVE value)   

Communication: USB port (Standard) and Blue tooth (Optional) 

Printer: Thermal Printer with blue tooth (optional) 

Power: 2 AA batteries 

Working temperature: -10

~ +50

 

Size: 153 × 76×37 (mm) (H×W×D) 

Weight: 280g include batteries 

Standard: GB/T 17394-2014, ASTM A956 

Warranty: 12 months after purchase 

3.2 Applications 

Installed machinery and Permanent assembly parts 

Mould cavity 

Heavy workpieces 

Failure analysis of pressure vessel, turbine and other equipments 

Small test area 

The production line of Bearings and other parts 

Distinguish the material of the metal material warehouse 

4 .Pre-Treatment of Workpiece 

4.1 Workpiece Requirements 

4.1.1 Surface Roughness requirements 

Surface roughness is an important requirement for the surface of the test piece, 

it should be smooth and no oil, or will cause measurement errors. 

The surface roughness requirements are listed in table 4.1 

Impact Device Type 

Work piece surface roughness Ra 

D, DC, D

15   

2um 

7um 

C   

0.4um 

4.1.2 Weight and Thickness Requirements 

Impact device 

type 

Min weight 

Min thickness

no coupling

 

Min thickness

coupling

 

D

DC

D

15 

5kg 

25mm 

3mm 

G   

15kg 

70mm 

10mm 

C   

1.5kg 

10mm 

1mm 

4.1.3 Surface Hardened Layer 

Impact device   

Min. depth of surface hardened layer

mm

 

D, DC, D

15   

0.8 

0.2 

 

4.1.4 Curved surface Requirement 

Curved surface: The best testing surface of sample is flat. When the curvature 

radius R of the surface to be tested is smaller than 30mm (D, DC, D+15, C, E 

and  DL  type  of  impact  device)  and  smaller  than  50mm  (G  type  of  impact 

device), the small support ring or the special support rings should be chosen. 

 

 

7                                                                                                                                                          8 

Summary of Contents for LM-100

Page 1: ...nts 8 4 1 2 Weight and Thickness Requirements 8 4 1 3 Surface Hardened Layer 8 4 1 4 Curved surface Requirement 8 4 2 Support of Test piece 9 5 Operation 9 5 1 Display 9 5 2 Keyboard 10 5 3 Turn on th...

Page 2: ...l voltage with the impact velocity The Leeb hardness values are calculated by the following formula HL 1000 Vb Va In Which HL Leeb hardness values Vb the voltage produced during the rebound of impact...

Page 3: ...n carbide Tungsten carbide Tungsten carbide Tungsten carbide Impact Device Diameter 20mm 20mm 20mm 30mm 20mm Impact Device Length 86 147 75mm 162mm 141mm 254mm 155mm Impact Device Weight 50g 80g 75g 2...

Page 4: ...uitcase 2 6 Optional accessories Impact device DC C DL D 15 G Standard Calibration Test Block Blue tooth module Thermal Printer Standard supporting ring Small Supporting ring Special supporting rings...

Page 5: ...s Mould cavity Heavy workpieces Failure analysis of pressure vessel turbine and other equipments Small test area The production line of Bearings and other parts Distinguish the material of the metal m...

Page 6: ...f a large weight and thickness may still cause deformation so reinforcement and support in the rear surface of the test point is necessary 5 Operation 5 1 Display 5 2 Keyboard 5 3 Turn on the instrume...

Page 7: ...irmly on the surface of the sample the impact direction should be vertical to the testing surface Press the release button on the upside of the impact device to test The sample and the impact device a...

Page 8: ...the Thermal Printer then the groups testing data will be printed 5 9 Calibration Error correction settings Measuring error can be usually caused by test tip abrasion or changing another impact device...