Calibrate
Section
6
YSI Incorporated
YSI 556 MPS
Page 45
7.
Stir the water with a stir bar, or by rapidly moving the probe
module, to provide fresh sample to the DO sensor.
8.
Allow at least one minute for temperature equilibration before
proceeding. The current values of all enabled sensors will
appear on the screen and will change with time as they
stabilize.
9.
Observe the DO mg/L reading, when the reading is stable
(shows no significant change for approximately 30 seconds),
press
Enter
. The screen will indicate that the calibration has
been accepted and prompt you to press
Enter
again to
Continue.
10.
Press
Enter
. This returns you to the DO calibration screen.
See Figure 6.7 DO Calibration Screen.
11.
Press
Escape
to return to the calibrate menu. See Figure 6.2
Calibrate Screen.
12.
Rinse the probe module and sensors in tap or purified water
and dry.
6.2.4
pH Calibration
1.
Go to the calibrate screen as described in
Section 6.2.1
Accessing the Calibrate Screen
.
2.
Use the arrow keys to highlight the
pH
selection. See Figure
6.2 Calibrate Screen.
3.
Press
Enter
. The pH calibration screen is displayed.