3.5.4. How do I set up an Echo Sounder in RiverSurveyor?
Before establishing communication with your echo sounder through the
RiverSurveyor
program:
•
Make sure the echo sounder is operating correctly (see echo sounder manual for details).
•
Make sure your echo sounder is using the appropriate communications settings. Typical se-
rial communication settings for an echo sounder are:
- Baud rate – 9600
- Data bits – 8
- Stop bits – 1
•
Keep in mind that both the ADP and echo sounder are acoustic instruments. Both emit and
receive similar types of signals. For this reason, there is a possibility of interference be-
tween the acoustic signals of the two instruments. Because RiverSurveyor is usually used in
shallow waters (less than 200 m), and because echo sounders use a single narrow beam,
simply increasing the physical distance between the two instruments will greatly minimize
the possibility of interference. For example, if the ADP is near the bow, place the echo
sounder near the stern.
You can now connect your echo sounder serial cable to your computer. Remember which serial
port you use, as you will need this information in the
RiverSurveyor
program.
The echo sounder input into the RiverSurveyor is set up via the
Systems
dialog box (
). The
Systems
dialog box can be displayed by selecting
Communications|Systems
or by
clicking the
Systems
Systems
dialog box.
In the
Systems
), select the serial (
Com
) port and
Baud Rate
for your echo
sounder. If the
Automatic Verification
is on (box checked), the
RiverSurveyor
program will auto-
matically search for a valid echo sounder data string (either SDDPT or SDDBT are acceptable)
as soon as the communications settings are entered. If
Automatic Verification
is off, click the
Click here to check
link in the
Status
column after entering your communications settings. If
valid echo sounder data are not found on the serial port, the
Status
column will change to
Invalid
– click here for information
. Click the link for information as to the possible reasons for the
problem (see Troubleshooting §
Figure 3-15.
Systems
dialog box in RiverSurveyor
(Echo Sounder)
Summary of Contents for SonTek RiverSurveyor
Page 2: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 ii ...
Page 4: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 iv ...
Page 14: ......
Page 24: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 12 ...
Page 26: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 14 ...
Page 74: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 62 ...
Page 82: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 70 ...
Page 148: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 136 ...
Page 162: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 150 ...
Page 166: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 154 ...
Page 168: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 156 ...
Page 182: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 170 ...
Page 194: ...SonTek YSI Inc RiverSurveyor System Manual May 1 2007 182 ...