40
9.
The
Var Max. Sample Volume
is displayed. Enter a value that
is larger than the Var Sample Volume 20 mA (see step #6) but
smaller than or equal to the maximum sample volume of the
containers being used (
e.g.
1000 mL when using 1000 mL
bottles).
10.
Highlight a
Var Function
and press the Enter key. This
determines what will happen if there is more solution that will
be placed in a bottle than the bottle can actually hold (see Var
Max. Sample Volume). It is recommended to highlight
Drop
Samples
and press the Enter key.
11.
The
Info
screen is displayed.
If there is only one bottle installed (composite sampling):
• Select
Settings Complete
and the ProSample will return to
the Programs menu. Skip to step #15.
If there is a distributor arm and multiple bottles installed
(discrete sampling):
• Select
More Settings
g
Bottle Selection
.
t
If you would like to fill all bottles in the sampler, select
All
Bottles
. Return to the More Settings menu by pressing the
Back key once, and then select
Programming OK
. Skip to
step #15.
t
If you want only to fill certain bottles, select
Bottle
Selection
and continue with step #12.
12. First Bottle
is displayed if the ProSample has more than one
bottle installed. Enter the bottle number in which the first
sample should be placed. Press the Enter key.
Programming: Analog Input Volume Modulation