YSI Pro2030 User Manual Download Page 16

24

25

up the value of the solution at the current temperature and enter that 

value into the Pro2030. Most conductivity solutions are labeled with a 

value at 25°C. If calibrating specific conductance, enter the value listed 

for 25°C. Depressing either the up or down arrow key for 5 seconds 

will move the changing digit one place to the left. The Pro2030 will 

remember the entered calibration value and display it the next time a 

conductivity calibration is performed.

Press  enter  to  complete  the  calibration.  or,  press  Cal  to  cancel  the 

4. 

calibration and return to the Run screen.    

‘Calibration  Successful’  will  display  for  a  few  seconds  to  indicate  a 

5. 

successful calibration and then the instrument will return to the Run 

screen.

If the calibration is unsuccessful, an error message will display on the 

6. 

screen.  Press  the  Cal  key  to  exit  the  calibration  error  message  and 

return to the Run screen.  See the Troubleshooting guide for possible 

solutions.  

CALIBRATING IN SALINITY

Fill a clean container (i.e. plastic cup or glass beaker) with fresh, traceable 

1. 

salinity  calibration  solution  and  place  the  sensor  into  the  solution. 

The solution must cover the holes of the conductivity sensor that are 

closest to the cable (figure 8). Ensure the entire conductivity sensor is 

submerged in the solution or the instrument will read approximately  

half the expected value. Gently move the probe up and down to remove 

any air bubbles from the conductivity sensor.

Turn the instrument on and allow the conductivity and temperature 

2. 

readings  to  stabilize.  Press  the  Cal  key.  Highlight  Conductivity  and 

press enter. Next, highlight Salinity and press enter. 

Use  the  up  or  down  arrow  key  to  adjust  the  value  on  the  display  to 

3. 

match the value of the salinity solution. Depressing either the up or 

down arrow key for 5 seconds will move the changing digit one place to 

the left. The Pro2030 will remember the entered calibration value and 

display it the next time a salinity calibration is performed.

Press  enter  to  complete  the  calibration.  or,  press  Cal  to  cancel  the 

4. 

calibration and return to the Run screen.    

‘Calibration  Successful’  will  display  for  a  few  seconds  to  indicate  a 

5. 

successful calibration and then the instrument will return to the Run 

screen.

If the calibration is unsuccessful, an error message will display on the 

6. 

screen.  Press  the  Cal  key  to  exit  the  calibration  error  message  and 

return to the Run screen. See the Troubleshooting guide for possible 

solutions.  

T A K I N G   M E A S U R E M E N T S

Before  taking  measurements,  be  sure  the  instrument  has  been  calibrated 

to  ensure  the  most  accurate  readings.  Turn  the  instrument  on  and  wait  5-15 

minutes if using a polarographic sensor. Install the sensor guard to protect the 

sensor and membrane. Place the probe in the sample to be measured and give the 

probe a quick shake to release any air bubbles. Be sure the conductivity sensor 

is completely submerged in the sample. The two holes near the cable should be 

covered by the sample for accurate conductivity readings (figure 8).  

Allow the temperature readings to stabilize. Next, stir the probe in the sample to 

overcome the stirring dependence of the dissolved oxygen sensor.  The dissolved 

oxygen sensor requires at least 6 inches (16 cm) per second of water movement 

if  using  the  yellow  membrane  and  3  inches  (7.62  cm)  per  second  of  water 

movement if using the blue membrane. The required sample movement can be 

achieved by the natural flow of the stream, physically stirring the probe in the 

sample or a combination of the two. once the values plateau and stabilize you 

may record the measurement and/or store the data set. The dissolved oxygen 

reading will drop over time if stirring or movement is ceased. If placing the Do 

sensor into a stream or fast flowing waters it is best to place it perpendicular to 

the flow and not facing into the flow.

If using the Do sensor near an aeration device, it is helpful to make sure air 

bubbles  do  not  burst  on  the  membrane  since  that  may  cause  unstable  Do 

readings. You should be able to prevent this by pointing the sensor upwards so 

it’s facing the sky and twist tying, zip tying or rubber banding the bulkhead to 

the cable. Essentially, making a simple curve to the cable without bending or 

breaking the cable will allow you to lower the sensor into the aerated sample 

while the sensor points skyward and air bubbles are no longer bursting on the 

membrane surface.

S AV I N G   A N D   V I E W I N G   D A T A

The Pro2030 can store 50 data sets in non-volatile memory for later viewing.  A 

data set includes the values currently on the display, i.e. temperature, dissolved 

oxygen and two conductivity parameters. Each data point is referenced with a 

data set number, 01 through 50.

savIng data

The  Pro2030  can  not  communicate  to  a  PC  via  a  Pro  Series 

communications saddle. Connecting the Pro2030 to a communication 

saddle may cause erratic instrument behavior. 

i

Summary of Contents for Pro2030

Page 1: ...USER MANUAL English Pro2030 ...

Page 2: ...Navigation 7 First Power On 8 System Setup Menu 8 DO Local 9 Last Digit Suppression LDS 10 Quick DO Calibration Quick DO Cal 10 Audio 11 Contrast 11 Sensor Type 11 Membrane Type 12 Auto Stable 13 DO Units 14 Conductivity Units Cond Units 14 Specific Conductance Reference Temperature SPC Ref Temp 15 Specific Conductance Temperature Coefficient SPC C 16 TDS Constant 16 Temperature Units 17 Pressure ...

Page 3: ...lace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact YSI Customer Service in Yellow Springs Ohio at 1 937 767 7241 800 897 4151 or visit www YSI com for a Product Return Form Send the product and proof of purchase transportation prepaid to the Authorized Service Center selec...

Page 4: ...parts with items on the packing list If any parts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased Battery Installation The instrument requires 2 alkaline C cell batteries Under normal conditions the average battery life is 425 hours at room temperature without using the back lig...

Page 5: ...turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Press and hold for 3 seconds to turn instrument off 4 Menu Use to enter the System Setup menu from the Run screen 5 Enter Press to confirm entries and selections 6 Down Arrow Use to navigate through menus to navigate through box options at the bottom of the Run screen and to decrease numerical inpu...

Page 6: ...head connector on the cable are 3 clean and dry Grasp the sensor with one hand and the cable bulkhead in the other 4 Push the sensor into the connector on the cable until it is firmly seated and 5 only 1 o ring is visible Failure to properly seat the sensor may result in damage Twist the sensor clockwise to engage the threads and finger tighten Do 6 NOT use a tool This connection is water tight Fo...

Page 7: ...information before displaying the main Run screen figure 5 The first time the Pro2030 is turned on it will step through language sensor and membrane selections see the First Power On section of this manual for more information Figure 5 Run screen Backlight Once the instrument is powered on pressing the power backlight key will turn on the display backlight The backlight will remain on until the ke...

Page 8: ...tup menu contains multiple screens that are notated as pages The current page is indicated near the bottom of the display figure 7 Use the up and down arrow keys to scroll through menu options and menu pages Exiting the System Setup Menu To exit the System Setup menu press the down arrow key until the ESC Exit box is highlighted then press enter to return to the Run screen DO Local DO Local can be...

Page 9: ... For more information on Quick DO Cal see the Calibration section of this manual Audio Audio can be enabled or disabled by using the up or down arrow keys to highlight Audio and pressing enter When enabled there will be an X in the box next to Audio When Audio is enabled the Pro2030 will beep twice to indicate stability when Auto Stable is enabled The instrument will also beep when a key is presse...

Page 10: ...itional Teflon membranes 5914 Blue 2 0 mil polyethylene PE Less flow dependence than 1 25 mil membrane but somewhat slower response Polarographic Membrane Kits Item Color Material Description 5908 Yellow 1 25 mil polyethylene PE Faster response time and less flow dependence than traditional Teflon membranes 5909 Blue 2 0 mil polyethylene PE Less flow dependence than 1 25 mil membrane but somewhat ...

Page 11: ... temperature values using algorithms found in Standard Methods for the Examination of Water and Wastewater TDS g L displays Total Dissolved Solids in grams per liter TDS is calculated from conductivity and temperature using a user selectable TDS constant TDS mg L displays Total Dissolved Solids in milligrams per liter TDS is calculated from conductivity and temperature using a user selectable TDS ...

Page 12: ...ess enter to open the submenu With the TDS Constant highlighted press enter to make the field adjustable Next use the up or down arrow key to increase or decrease the value The input range is 0 30 to 1 00 Press enter to save the new TDS Constant Next highlight the ESC Exit box and press enter to close the submenu Temperature Units Highlight Temperature Units and press enter to open a submenu that ...

Page 13: ... the Pro2030 is calibrated at the factory The barometer reading must be accurate to ensure accurate DO calibrations and readings If the barometer requires an adjustment use the up or down arrow keys to highlight the barometer box along the bottom of the Run screen then press enter Next use the up or down arrow keys to adjust the barometer reading to the local true barometric pressure Continually d...

Page 14: ...necessary to calibrate in both and mg L or ppm Calibrating in will simultaneously calibrate mg L and ppm and vice versa YSI recommends calibrating the dissolved oxygen sensor in for both ease and accuracy i Calibrating in Percent DO Perform this calibration procedure when Quick DO Cal is disabled in the System Setup menu Perform steps 1 4 of the Quick DO Calibration procedure 1 Press and hold the ...

Page 15: ...e Reference Temperature and the SpecificConductanceTemperatureCoefficientregardlessofwhatisconfiguredin the System Setup Menu The default value for the Reference Temperature is 25 C It is not necessary to calibrate conductivity specific conductance and salinity Calibrating one of these parameters will simultaneously calibrate the others YSI recommends calibrating specific conductance for greatest ...

Page 16: ...ey to exit the calibration error message and return to the Run screen See the Troubleshooting guide for possible solutions T aking M easurements Before taking measurements be sure the instrument has been calibrated to ensure the most accurate readings Turn the instrument on and wait 5 15 minutes if using a polarographic sensor Install the sensor guard to protect the sensor and membrane Place the p...

Page 17: ...to access Data mode Note that the function boxes at the bottom of the display are different in Data mode figure 10 Figure 10 Data mode Viewing Data Once in Data mode use the up and down arrow keys to view saved data sets in sequential order or press enter to access the bottom functions After accessing the bottom functions highlight the Data box and press enter to regain access to viewing data The ...

Page 18: ...ng properly If the gasket o rings and sealing surfaces are not maintained properly it is possible that water can enter the battery compartment and or sensor port of the instrument If water enters these areas it can severely damage the battery terminalsorsensorportcausinglossofbatterypower falsereadingsandcorrosion to the sensors or battery terminals Therefore when the battery compartment lid is re...

Page 19: ...over grease the o rings The excess grease may collect grit particles that can compromise the seal Excess grease can also cause the waterproofing capabilities of the o ring to diminish potentially causing leaks If excess grease is present remove it using a lens cloth or lint free cloth i i Sensor Maintenance Conductivity The openings that allow sample access to the conductivity electrodes should be...

Page 20: ...sensor tip directly down on the sanding disc and twist it in a circular motion to sand the gold cathode The goal is to sand off any build up and to lightly scratch the cathode to provide a larger surface area for the electrolyte solution under the membrane Usually 3 to 4 twists of the sensor are sufficient to remove deposits and for the gold to appear to have a matte finish Rinse the sensor thorou...

Page 21: ...ranes Chemical cleaning should be performed as infrequently as possible First attempt a membrane change and recalibrate If a new membrane does not resolve the problem then proceed with cleaning i At the time the cap membrane is changed YSI recommends that you rinse the anode and cathode figure 12 with distilled or deionized water and wipe with a clean paper towel If white deposits are evident on t...

Page 22: ... membrane 2 type selection in the System Setup menu Calibration sleeve may not be 3 100 water saturated ensure sponge is moistened Ensure adequate sample movement if 4 performing mg L or ppm calibration Allow sufficient stabilization time for 5 dissolved oxygen and temperature AND wait at least 3 seconds before confirming a DO or DO Local calibration Replace membrane and electrolyte 6 Clean sensor...

Page 23: ...y calibration 3 Verify temperature readings 4 are accurate Sample conductivity is outside 5 the measurement range of the instrument i e 0 200 mS Contact YSI Tech Support 6 S pecificati o ns These specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at www ysi com or contact YSI Tech Support...

Page 24: ... a temperature and conductivity sensor The dissolved oxygen sensor is sold separately D eclarati o n o f C o nf o rmity The undersigned hereby declares on behalf of the named manufacturer under our sole responsibility that the listed product conforms to the requirements for the listed European Council Directive s and carries the CE mark accordingly Manufacturer YSI Incorporated 1725 Brannum Lane Y...

Page 25: ... and users are expected to understand and follow the battery disposal requirements for their specific locale C o ntact I nf o rmati o n Ordering and Technical Support Telephone 800 897 4151 USA 1 937 767 7241 Globally Monday through Friday 8 00 AM to 5 00 ET Fax 1 937 767 9353 orders 1 937 767 1058 technical support Email environmental ysi com Mail YSI Incorporated 1725 Brannum Lane Yellow Springs...

Page 26: ... Air at 760 mm Hg Pressure Salinity Measure of quantity of dissolved salts in water Chlorinity Measure of chloride content by mass of water S 0 00 1 80655 x Chlorinity 0 00 Temp C Chlorinity 0 Salinity 0 5 0 ppt 9 0 ppt 10 0 ppt 18 1 ppt 15 0 ppt 27 1 ppt 20 0 ppt 36 1 ppt 25 0 ppt 45 2 ppt 0 0 14 62 13 73 12 89 12 10 11 36 10 66 1 0 14 22 13 36 12 55 11 78 11 07 10 39 2 0 13 83 13 00 12 22 11 48 ...

Page 27: ...7 6 73 6 41 6 10 5 81 32 0 7 31 6 96 6 62 6 31 6 01 5 72 33 0 7 18 6 84 6 52 6 21 5 91 5 63 34 0 7 07 6 73 6 42 6 11 5 82 5 55 35 0 6 95 6 62 6 31 6 02 5 73 5 46 36 0 6 84 6 52 6 22 5 93 5 65 5 38 37 0 6 73 6 42 6 12 5 84 5 56 5 31 38 0 6 62 6 32 6 03 5 75 5 48 5 23 39 0 6 52 6 22 5 98 5 66 5 40 5 15 40 0 6 41 6 12 5 84 5 58 5 32 5 08 41 0 6 31 6 03 5 75 5 49 5 24 5 01 42 0 6 21 5 93 5 67 5 41 5 1...

Reviews: