28
air. If the connector is corroded, contact YSI Technical Support or the
Authorized dealer where you purchased the instrument.
Remove sensors upside down (facing the ground) to help
prevent water from entering the port upon removal.
i
SENSOR MAINTENANCE
Sensor Maintenance - Temperature
You must keep the temperature portion of the sensor free of build up.
Other than that, the sensor requires no maintenance. A toothbrush can
be used to scrub the temperature sensor if needed.
Sensor Maintenance - Dissolved Oxygen
Membrane Cap Installation
The DO sensor (Polarographic and Galvanic) is shipped with a dry,
protective red cap that will need to be removed before using. Remove
the protective cap or used membrane cap and replace it with a new
membrane cap following these instructions:
Remove the sensor guard to access the sensor.
Remove the protective red cap or unscrew and remove
any old membrane cap by holding the sensor when
unscrewing the membrane cap and discard.
Thoroughly rinse the sensor tip with distilled or
deionized water.
Fill a new membrane cap with O
2
sensor
solution that has been prepared according
to the directions on the bottle. Be very
careful not to touch the membrane surface.
Lightly tap the side of the membrane cap to
release bubbles that may be trapped.