![YSI Pro10 User Manual Download Page 9](http://html.mh-extra.com/html/ysi/pro10/pro10_user-manual_915322009.webp)
10
11
you have confirmed any changes, highlight the ESC-Exit box along
the bottom of the display and press enter to close the Auto Stable
submenu. To disable Auto Stable, set the % Change input to 0.0.
ph BUffer set
Highlight pH Buffer Set and press enter to open a submenu that allows
you to select the Buffer Set used for auto buffer recognition during a
pH calibration. There are two buffer set options: USA (4, 7 and 10) and
NIST (4.01, 6.86 and 9.18). Highlight the buffer set and press enter to
enable. The enabled buffer set will have an ‘X’ in the box next to it.
Highlight the ESC-Exit box and press enter to save any changes and to
close the submenu.
langUage
Highlight Language and press enter to open a submenu that allows
you to change the language. Highlight the desired language (English,
Spanish, Portuguese, or French) and press enter to enable. The enabled
language will have an ‘X’ in the box next to it. Highlight the ESC-Exit
box and press enter to save any changes and to close the Language
submenu.
The text in the boxes along the bottom of the Run screen will always be
displayed in English regardless of the language enabled in the System
Setup menu.
teMperatUre UnIts
Highlight Temperature Units and press enter to open a submenu that
allows you to change the temperature units displayed on the Run
screen. Highlight the desired unit (Celsius or Fahrenheit) and press
enter to enable. The enabled temperature unit will have an ‘X’ in the
box next to it. Only one unit may be enabled at a time. Highlight the
ESC-Exit box and press enter to save any changes and to close the
Temperature Units submenu.
aUto shUtoff
Auto Shutoff allows you to set the instrument to turn off automatically
after a period of time since the last button press. In the setup menu, use
the up or down arrow keys to highlight Auto Shutoff, then press enter
to open the submenu. Press enter while the minute field is highlighted
to make it adjustable. Next, use the up or down arrow keys to adjust the
shut off time from 0 to 60 minutes. Press enter to save the new shutoff
time. Next, highlight the ESC-Exit box and press enter to close the
submenu. To disable Auto Shutoff, set the Time in Minutes to 0 (zero).
resettIng the systeM setUp MenU
to factory defaUlt
To reset the Pro10 settings back to factory default, press the down
arrow key while in the System Setup menu until the Reset - box
is highlighted, then press enter. The instrument will prompt you to
confirm the reset. Highlight Yes and press enter to continue with the
reset or highlight No and press enter to cancel the reset. A Factory
Reset will not affect data saved in the instrument’s memory.
The following will be set in the Pro10 after performing a reset:
Parameter
Reset Defaults
Audio
On
Contrast
Set to mid range
ISE Sensor Type
pH
ISE Units
pH
ISE Auto Stable
Off (0.0 % Change
and 10 seconds)
pH Buffer Set
USA
Temperature Units
°C
Language
English
Auto Shutoff
30 minutes
ISE Calibration
Reset to factory default*
*It is recommended to perform an ISE calibration after performing a
reset.
C A L I B R A T I O N
teMperatUre
All Pro10 cables have built-in temperature sensors. Temperature
calibration is not required nor is it available.
ph calIBratIon
The Pro10 pH sensors can be calibrated by performing a 1, 2 or 3-point
calibration. At least one of the calibration points must be done with