YSI Pro 1020 Operating Procedure Download Page 7

IOP 019B.0 

Effective Date: 1/26/2017 

Page 7 of 8 

 

2. 

pH sensor - The key to pH sensor storage, short or long-term, is to make certain 

that the sensor does not dry out. Sensors which have been allowed to dry out due 

to improper storage procedures may be irreparably damaged by the dehydration 

and will require replacement. You can try to rehydrate the sensor by soaking it 

(preferably overnight) in a potassium chloride solution or a pH 4 buffer before 

attempting to calibrate. To store the sensor, remove it from the cable and seal the 

vacant port with a port plug. Fill the original shipping/storage vessel (plastic 

boot or bottle) with buffer 4 solution and then submerge the sensor into the 

solution. The sensor should remain submerged in the solution during the storage 

period; therefore, make certain that the vessel is sealed to prevent evaporation 

and periodically check the vessel to ensure the sensor does not dry out. 

MAINTENANCE: 

 

I. 

Routine cleaning of pH sensor at the end of each treatment session or when pH response is slow 

A. 

Use clean water and a soft clean cloth, lens cleaning tissue, or cotton swab to remove all 

foreign material from the glass bulb or platinum button. Then use a moistened cotton 

swab to carefully remove any material that may be blocking the reference electrode 

junction of the sensor. 

B. 

Soak the sensor for 10-15 minutes in clean water (Not DI water. Soaking in DI water will 

shorten the life span of the probe) containing a few drops of commercial dish washing 

liquid. Gently clean the glass bulb and platinum button by rubbing with a cotton swab 

soaked in the cleaning solution. Rinse the sensor in clean water, wipe with a cotton swab 

saturated with clean water, and then re-rinse with clean water. 

II. 

Additional cleaning when good pH response is not restored using Step I 

A.

 

Soak the sensor for 30-60 minutes in one molar hydrochloric acid. Rinse the sensor in clean 

water, wipe with a cotton swab saturated with clean water, and then re-rinse with clean water. 

To be certain all traces of acid are removed, soak the sensor in clean water (Not DI water. 

Soaking in DI water will shorten the life span of the probe) for about an hour with occasional 

stirring. 

B. 

If biological contamination of the reference junction is suspected or good response is not 

restored, soak the sensor for approximately 1 hour in a 1:1 dilution of household chlorine 

bleach. Rinse the sensor with clean water and then soak for at least 1 hour (more if 

possible) in clean water (Not DI water. Soaking in DI water will shorten the life span of 

the probe) with occasional stirring to remove residual bleach from the junction. Then re-

rinse the sensor with clean water and retest. 

III. 

For additional troubleshooting, see user manual. 

REFERENCE: 

 

User Manual for YSI Pro 1020 

Quick Start Guide for YSI Pro 1020 

Calibration Tips for YSI Professional Plus Multiparameter Water Quality Meter 

 

 

 

 

 

Summary of Contents for Pro 1020

Page 1: ...bes pH 605101 DO 605203 MANUFACTURER Yellow Springs Instrument Co Inc YSI PRECAUTIONS POTENTIAL INTERFERENCES General Greatest accuracy is achieved if the actual samples to be measured are within 10 C of the calibration solutions Longer cable lengths introduce more error into the readings Dissolved oxygen Chlorine sulfur dioxide nitric oxide and nitrous oxide can affect readings by behaving like o...

Page 2: ...nsor and two sensor ports 3 Sensors two sensors are required to measure all of the water chemistry parameters indicated above Sensors are installed in the cable bulkhead ports indicated below to ensure proper function If measurement of one of the parameters is not required the corresponding sensor s may be removed and replaced with a sensor port plug a temperature sensor is permanently incorporate...

Page 3: ...ess the Power key to turn the instrument on The instrument will briefly display the splash screen with the YSI logo then go directly to the main run screen 2 Setup Sensors a Press Menu button b Dissolved oxygen sensor 1 Highlight DO Sensor Type Press enter 2 Highlight Polaro black and press enter to confirm 3 Highlight ESC exit press enter 4 Highlight DO Membrane type press enter 5 Highlight 1 25 ...

Page 4: ...lt in damage e Twist the sensor clockwise to engage threads and finger tighten Do not use a tool This connection is waterproof Refer to the sensor installation sheet included with each sensor for detailed instructions 5 Install the 2003 DO sensor membrane as follows a Change membrane and oxygen probe electrolyte solution every 30 days b Remove the metal sensor guard 3 If installing a membrane for ...

Page 5: ...reading pH mVs in buffer 4 should be 159 to 180 mV from the previous buffer 7 pH mV value pH mVs in buffer 10 should be 159 to 180 mV from the previous buffer 7 pH mV value If the mV readings are outside of these ranges clean the sensor as described in the maintenance section of this protocol 10 Press enter to complete the calibration or press Cal to cancel 11 Calibration Successful will display f...

Page 6: ...nd across the DO sensor membrane to obtain an accurate reading Be careful not to create a vortex which allows air bubbles to pass against the DO membrane G After at least 5 minutes and once the values stabilize record the measurement IIX Sensor storage A Short term storage The cable assembly is supplied with a sensor storage container or sleeve that attaches to the cable The container is used for ...

Page 7: ... 15 minutes in clean water Not DI water Soaking in DI water will shorten the life span of the probe containing a few drops of commercial dish washing liquid Gently clean the glass bulb and platinum button by rubbing with a cotton swab soaked in the cleaning solution Rinse the sensor in clean water wipe with a cotton swab saturated with clean water and then re rinse with clean water II Additional c...

Page 8: ...Jessica M Barber Digitally signed by JESSICA BARBER Date 2020 03 10 11 25 00 04 00 ...

Reviews: