YSI Pro 1020 Operating Procedure Download Page 6

IOP 019B.0 

Effective Date: 1/26/2017 

Page 6 of 8 

 

completely saturated and to allow the sensors to stabilize. 

3.

 

Press and hold 

Cal 

for 3 seconds. Highlight 

Dissolved Oxygen

 and press enter 

to confirm. 

4.

 

The meter will indicate 

Calibrating %DO 

 on the display 

5.

 

The instrument will use the internal barometer during calibration and will 

display this value in the bar at the bottom of the display. 

6.

 

Calibration Successful will display for a few seconds to indicate a successful 

calibration and then the instrument will return to the Run screen. 

7.

 

If the Calibration is unsuccessful, an error message will display on the screen.  

Press the Cal key to exit the error message and to return to the Run screen.  See 

Troubleshooting section of the user manual for possible solutions. 

VII. 

Measurements 

A. 

Ensure the cable and sensors are tightly connected to the meter. 

B. 

Be sure the meter has been calibrated. 

C. 

Place the sensors in the sample to be measured. 

D. 

Shake the probe in the sample to release air bubbles. 

E. 

Allow temperature reading to stabilize. 

F. 

Stir the sample water with the probe at a moderate pace. You must provide 6 inches of 

water movement per second across the DO sensor membrane to obtain an accurate 

reading. Be careful not to create a vortex which allows air bubbles to pass against the DO 

membrane. 

G. 

After at least 5 minutes and once the values stabilize, record the measurement. 

IIX. 

Sensor storage 

 A. 

Short term storage 

The cable assembly is supplied with a sensor storage container, or sleeve, that attaches to 

the cable. The container is used for short-term storage (less than 7 days). Be sure to keep 

a small amount of moisture (tap water) in the container during storage. This is done to 

maintain a 100% saturated air environment which is ideal for short-term sensor storage. 

The sensors should not be submersed in water. The intent is to create a humid air storage 

environment. 

B. 

Long term storage 

1. 

Dissolved oxygen sensor - Dissolved oxygen sensors should be stored in a dry 

state for long term storage. First, remove the membrane cap and thoroughly 

rinse the sensor with clean water. Next, either blow it dry with compressed air or 

allow to air dry completely. Install a clean, dry new membrane cap over the 

sensor to keep it dry and to protect the electrodes. After storing the sensor for a 

long period of time, it is necessary to “condition” the sensor by putting a new 

membrane with electrolyte solution on the sensor and then turning the 

instrument on to allow the sensor sufficient time to stabilize. 

Summary of Contents for Pro 1020

Page 1: ...bes pH 605101 DO 605203 MANUFACTURER Yellow Springs Instrument Co Inc YSI PRECAUTIONS POTENTIAL INTERFERENCES General Greatest accuracy is achieved if the actual samples to be measured are within 10 C of the calibration solutions Longer cable lengths introduce more error into the readings Dissolved oxygen Chlorine sulfur dioxide nitric oxide and nitrous oxide can affect readings by behaving like o...

Page 2: ...nsor and two sensor ports 3 Sensors two sensors are required to measure all of the water chemistry parameters indicated above Sensors are installed in the cable bulkhead ports indicated below to ensure proper function If measurement of one of the parameters is not required the corresponding sensor s may be removed and replaced with a sensor port plug a temperature sensor is permanently incorporate...

Page 3: ...ess the Power key to turn the instrument on The instrument will briefly display the splash screen with the YSI logo then go directly to the main run screen 2 Setup Sensors a Press Menu button b Dissolved oxygen sensor 1 Highlight DO Sensor Type Press enter 2 Highlight Polaro black and press enter to confirm 3 Highlight ESC exit press enter 4 Highlight DO Membrane type press enter 5 Highlight 1 25 ...

Page 4: ...lt in damage e Twist the sensor clockwise to engage threads and finger tighten Do not use a tool This connection is waterproof Refer to the sensor installation sheet included with each sensor for detailed instructions 5 Install the 2003 DO sensor membrane as follows a Change membrane and oxygen probe electrolyte solution every 30 days b Remove the metal sensor guard 3 If installing a membrane for ...

Page 5: ...reading pH mVs in buffer 4 should be 159 to 180 mV from the previous buffer 7 pH mV value pH mVs in buffer 10 should be 159 to 180 mV from the previous buffer 7 pH mV value If the mV readings are outside of these ranges clean the sensor as described in the maintenance section of this protocol 10 Press enter to complete the calibration or press Cal to cancel 11 Calibration Successful will display f...

Page 6: ...nd across the DO sensor membrane to obtain an accurate reading Be careful not to create a vortex which allows air bubbles to pass against the DO membrane G After at least 5 minutes and once the values stabilize record the measurement IIX Sensor storage A Short term storage The cable assembly is supplied with a sensor storage container or sleeve that attaches to the cable The container is used for ...

Page 7: ... 15 minutes in clean water Not DI water Soaking in DI water will shorten the life span of the probe containing a few drops of commercial dish washing liquid Gently clean the glass bulb and platinum button by rubbing with a cotton swab soaked in the cleaning solution Rinse the sensor in clean water wipe with a cotton swab saturated with clean water and then re rinse with clean water II Additional c...

Page 8: ...Jessica M Barber Digitally signed by JESSICA BARBER Date 2020 03 10 11 25 00 04 00 ...

Reviews: