INSTRUMENT OPERATING PROCEDURE 

INSTRUMENT:

 

Multiparameter Water Quality Instrument 

MODEL: 

-

 

Pro 1020 Meter (6051020) 

-

 

Cable (6051020) 

-

 

Probes: pH (605101), DO (605203) 

MANUFACTURER:

 

Yellow Springs Instrument Co., Inc. (YSI) 

PRECAUTIONS: 

POTENTIAL INTERFERENCES 

General 

 –  

-

 

Greatest accuracy is achieved if the actual samples to be measured are within 10 ˚C of the 

calibration solutions. 

-

 

Longer cable lengths introduce more error into the readings. 

Dissolved oxygen

 – 

-

 

Chlorine, sulfur dioxide, nitric oxide, and nitrous oxide can affect readings by behaving like 

oxygen at the membrane. 

-

 

Static electricity may affect readings. 

-

 

Gold Cathode - The gold cathode on the sensor must be textured properly.  It may become 

tarnished or plated with silver after extended periods of use. 

-

 

Bubbles under the membrane cap will affect readings. 

-

 

Excessive water in calibration/storage sleeve will affect calibration. 

Summary of Contents for Pro 1020

Page 1: ...bes pH 605101 DO 605203 MANUFACTURER Yellow Springs Instrument Co Inc YSI PRECAUTIONS POTENTIAL INTERFERENCES General Greatest accuracy is achieved if the actual samples to be measured are within 10 C of the calibration solutions Longer cable lengths introduce more error into the readings Dissolved oxygen Chlorine sulfur dioxide nitric oxide and nitrous oxide can affect readings by behaving like o...

Page 2: ...nsor and two sensor ports 3 Sensors two sensors are required to measure all of the water chemistry parameters indicated above Sensors are installed in the cable bulkhead ports indicated below to ensure proper function If measurement of one of the parameters is not required the corresponding sensor s may be removed and replaced with a sensor port plug a temperature sensor is permanently incorporate...

Page 3: ...ess the Power key to turn the instrument on The instrument will briefly display the splash screen with the YSI logo then go directly to the main run screen 2 Setup Sensors a Press Menu button b Dissolved oxygen sensor 1 Highlight DO Sensor Type Press enter 2 Highlight Polaro black and press enter to confirm 3 Highlight ESC exit press enter 4 Highlight DO Membrane type press enter 5 Highlight 1 25 ...

Page 4: ...lt in damage e Twist the sensor clockwise to engage threads and finger tighten Do not use a tool This connection is waterproof Refer to the sensor installation sheet included with each sensor for detailed instructions 5 Install the 2003 DO sensor membrane as follows a Change membrane and oxygen probe electrolyte solution every 30 days b Remove the metal sensor guard 3 If installing a membrane for ...

Page 5: ...reading pH mVs in buffer 4 should be 159 to 180 mV from the previous buffer 7 pH mV value pH mVs in buffer 10 should be 159 to 180 mV from the previous buffer 7 pH mV value If the mV readings are outside of these ranges clean the sensor as described in the maintenance section of this protocol 10 Press enter to complete the calibration or press Cal to cancel 11 Calibration Successful will display f...

Page 6: ...nd across the DO sensor membrane to obtain an accurate reading Be careful not to create a vortex which allows air bubbles to pass against the DO membrane G After at least 5 minutes and once the values stabilize record the measurement IIX Sensor storage A Short term storage The cable assembly is supplied with a sensor storage container or sleeve that attaches to the cable The container is used for ...

Page 7: ... 15 minutes in clean water Not DI water Soaking in DI water will shorten the life span of the probe containing a few drops of commercial dish washing liquid Gently clean the glass bulb and platinum button by rubbing with a cotton swab soaked in the cleaning solution Rinse the sensor in clean water wipe with a cotton swab saturated with clean water and then re rinse with clean water II Additional c...