MultiLab 4010-1W
Settings
ba76141e07
08/2018
69
9.3.2
Enter
Sensor cap coefficients
(ProOBOD)
9.3.3
DO % Saturation local
Irrespective of the height or air pressure, the calibration value is set to 100 %.
The function
DO % Saturation local
fulfills the EU regulations for the parameter
oxygen saturation [%].
When the
DO % Saturation local
is enabled the display shows an [L] for the
parameter oxygen saturation.
D.O. mg/L readings are unaffected by the selection of the
DO % Saturation
local
function.
Resolution
(only for ProOBOD))
0.1
1
Set a high or low resolution.
The setting of the resolution is sep-
arately stored for each measured
parameter.
Response time t90
(only for FDO 4410)
30
... 300 s
Response time of the signal filter (in
seconds).
A signal filter in the sensor reduces
the limits of variation of the mea-
sured value. The signal filter is
characterized by the response time
t90. This is the time after which 90
% of a signal change is displayed.
DO % Saturation
local
on
off
DO % Saturation local
is a proce-
dure that takes the local air pres-
sure into account for each
saturation measurement.
For details, see section 9.3.3 DO %
S
Reset
-
Resets all sensor settings to the
delivery condition (see section
9.6.1 R
The values of the coefficients are provided with the sensor cap.
1.
Change the digit of the highlighted position with
<
><
>
.
2.
Go to the next position with
<
><F2>/[
]
.
3.
Confirm with
<ENTER>
when a coefficient is completely entered.
Menu item
Possible
setting
Explanation