MultiLab 4010-1W
Settings
ba76141e07
08/2018
75
9.5.5
Display illumination
The meter automatically switches off the display illumination if no key is
pressed for 20 seconds.
The illumination is switched on with the next keystroke again.
You can also generally switch on the display illumination (see section 9.5.1
S
9.6
Reset
You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.
9.6.1
Resetting the measurement settings
pH
The following settings for pH measurements are reset to the default settings
with the
Reset
function:
The sensor settings are reset under the
Reset
menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
<ENTER>
key.
The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Buffer
YSI
Calibration interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
Unit for slope
0.001
mV resolution
0.1
Asymmetry
0 mV
Slope
-59.2 mV
Man. temperature
25 °C
Single-point calibration
off