pH Calibration 

The pH100A uses a 2-point calibration. The first point must be a 6.86/7.00 buffer, and the 
second either a 4.00/4.01 or 9.18/10.01. These buffers can be purchased from a YSI 
representative.  
1.  Turn the unit on. Connect the pH electrode to the BNC connector and the ATC/Temp 

probe to the ATC/Temp connector of the unit; “ATC” displays. Press MODE until “pH” 
displays.  Autolock may be on or off as desired. 

2.  Place the pH and ATC/temp probes into the first buffer solution (either 7.00 or 6.86). 

Allow temperature readings to stabilize, then press and hold

 

“STAND” for 3 seconds

 

to 

calibrate.  If AUTOLOCK is off, the first point has been calibrated. If AUTOLOCK is on, 
“WAIT” flashes until the unit detects a stable reading. Once the unit calibrates the first 
point, “SLOPE” flashes. NOTE: If no temperature probe is connected, adjust the 
temperature reading to that of the first buffer using the 

 or 

 keys (0.0 to 60ºC) BEFORE 

pressing “STAND”. 

3.   Rinse the pH and ATC/temp probes in distilled water, then place into the second buffer 

solution (either 4.01/4.00 or 10.01/9.18). Allow temperature readings to stabilize, then 
press “SLOPE” to calibrate.  If AUTOLOCK is off, the second point has been calibrated. If 
AUTOLOCK is on, “WAIT” flashes until the unit detects a stable reading. Once the unit 
calibrates the second point, the unit beeps twice and

 

both “STAND” and “SLOPE” display 

steadily.  NOTE: If no temperature probe is connected, adjust the temperature reading to 
that of the first buffer using the 

 or 

 keys (0.0 to 60ºC) BEFORE pressing “SLOPE”. 

4.  The unit calculates and compensates for the pH electrode slope deviation corresponding 

to the values of the two calibration buffers. The unit is now dual-point calibrated and 
ready for measurements. After calibration, press and hold MEA. /EFF. for about 5 
seconds to display the new electrode efficiency. 

 

Using the model pH160 Electrode Simulator 

The model pH160 Electrode Simulator can be used to confirm proper instrument calibration. 
To use the simulator: 
1.  Install the 9V battery provided. 
2.  Attach the pH160 to the pH100A. Turn both units on. The pH160 has a small switch to the 

right of the pH buttons.   

3.  In pH measurement mode, press one of the pH buffer simulator buttons on the pH160. 

The corresponding pH value should appear on both screens. 

Note: Calibration with the pH simulator calibrates only the instrument - NOT the instrument 
and probe. For best accuracy, calibrate the pH instrument and probe together using buffer 
solutions. 
 

pH Measurements 

To take pH measurements, “STAND” and “SLOPE” must display steadily, indicating the unit is 
dual-point calibrated and ready for measurements.  If “STAND” and “SLOPE” are blinking, 
perform a pH calibration before taking measurements.  
1. Press 

MODE to enter pH mode with AUTOLOCK on or off as desired. For inherently 

unstable samples, the unit will not AUTOLOCK. Turn AUTOLOCK off in this case. 

2.  Rinse the pH electrode and/or ATC/temp probe with distilled water and immerse in the 

sample to be measured. Remove any air bubbles trapped around the probe by shaking or 
stirring the probe. Allow the pH and/or temperature to stabilize. If no ATC/temp probe is 
connected, “MAN” displays, indicating manual temperature compensation. Set unit to 
display the sample temperature by pressing the 

 and 

 keys (-10.0 to 120°C). If an 

ATC/temp probe is connected “ATC” displays along with the sample temperature.   

3.  If AUTOLOCK is off, the pH value of the sample displays on the screen. If both pH and 

temperature readings are stable, take a reading. If AUTOLOCK is on, press MEA./EFF.  
“WAIT” flashes until the unit determines a stable pH reading. 

 

Summary of Contents for EcoSense pH100A

Page 1: ...Operations Manual EcoSense pH100A Portable pH mV and Temperature Instrument English ...

Page 2: ...isposal 3 KEY FUNCTIONS OF THE MODEL pH100A 4 THE LCD DISPLAY 4 OPERATIONAL PROCEDURES 4 Buffer Set Selection 4 pH Calibration 5 Using the model pH160 Electrode Simulator 5 pH Measurements 5 Temperature Measurements 6 mV Measurements 6 Saving Viewing and Deleting Data 6 TROUBLESHOOTING 7 SPECIFICATIONS 7 ...

Page 3: ...et memory and 50 60 Hz AC noise rejection This meter is universal user friendly for field industrial and laboratory applications INITIAL INSPECTION Carefully unpack the unit and accessories and inspect for shipping damages Compare received parts with materials listed on the packing list Notify YSI immediately of any damage or missing parts Save all packing materials until satisfactory operation is...

Page 4: ...ternal memory When ESC is pressed for 2 seconds all LCD elements light After about 2 seconds the unit enters pH AUTOLOCK mode AUTOLOCK displays and STAND begins to flash indicating the need for calibration THE LCD DISPLAY 1 WAIT Displays while unit waits for a stable reading or end point sensing 2 BAT Low battery indicator 3 ATC MAN ATC displays if an ATC probe is connected Otherwise MAN displays ...

Page 5: ... calibrated and ready for measurements After calibration press and hold MEA EFF for about 5 seconds to display the new electrode efficiency Using the model pH160 Electrode Simulator The model pH160 Electrode Simulator can be used to confirm proper instrument calibration To use the simulator 1 Install the 9V battery provided 2 Attach the pH160 to the pH100A Turn both units on The pH160 has a small ...

Page 6: ...SAVE and the record number for one second Full is displayed when trying to save data and the memory is full To view saved data press mode until RECALL is displayed and then press ESC Use the Up or Down arrow keys to review different saved records Press Mode to escape back to measurement mode To delete data records press Mode while in measurement mode until DELETE is displayed Press ESC All will be...

Page 7: ...r temperature within range Return for service OvEr Undr Measured pH is out of the 16 00 2 00 pH range Measured mV is out of the 1250 2000 mV range Measured temperature is out of the 10 120 C range Bring the out of range unit into the correct measuring range If units are within proper range return product for service SPECIFICATIONS Display Range Accuracy Resolution pH 2 00 to 16 00 pH 0 1 2 lsd 0 0...

Reviews: