YSI EcoSense DO200A User Manual Download Page 13

9

6.  When the proper pressure displays, press Enter 

 once to view the 

calibration value in the lower right of the display. Once the value in the 
main display stabilizes, press Enter 

 again to move to the salinity 

compensation procedure.

7.  The display prompts for the approximate salinity of the water to be 

analyzed. Use the Up and Down keys to increase or decrease the 
salinity compensation value to the value of your sample (between 0 to 
40 parts per thousand [ppt]). When the correct salinity displays, press 
Enter 

.

8.  The unit holds calibration even if it is powered off. However, it is 

recommended to check calibration with each use and recalibrate 
as necessary to prevent drift. Dissolved oxygen readings are only as 
good as the calibration. 

Saving, Viewing and Deleting Data

The DO200A can save 50 data records, while the DO200M can save 250 
data records. When in measurement mode, press Enter 

 to save a 

record. The instrument will confirm saving the data by displaying “Save” 
and the data record number for one second. “Full” is displayed when 
trying to save data and memory is full. 

To view saved data, press Mode until “Recall” is displayed and then press 
Enter 

. Use the Up or Down arrow keys to review different saved 

records. Press Mode to escape back to measurement mode. 

To delete data records, press Mode while in measurement mode until 
“Delete” is displayed. Press Enter 

. “All” will be displayed and blinking. 

Press the Up or Down arrow key to switch between delete ‘All’ or ‘Each’ 
options. Select either ‘All’ or ‘Each’ by pressing Enter 

 while that option 

is displayed. 

If ‘All’ is selected, all records will be deleted from memory and ‘None’ will 
be displayed. Press Mode twice to return to the measurement mode.

If ‘Each’ is selected, the Up and Down arrow keys will allow you to scroll 
through the saved data records. Press Enter to delete the selected record. 
All records after the deleted record will shift up to keep the records in 
sequential order. For example, if record 3 is deleted, record 4 will become 
record 3 and record 5 will become record 4. Press Mode twice to return to 
the measurement mode.    

606035REF-B-YSI-DO200A-DO200M-User-Manual-English.indd   9

12/7/2017   9:45:11 AM

Summary of Contents for EcoSense DO200A

Page 1: ...USER MANUAL English Fran ais Espa ol Portugu s EcoSense DO200A EcoSense DO200M Dissolved oxygen temperature instruments User Manual DOCUMENT 606035REF...

Page 2: ...amage or failure caused by i failure to install operate or use the product in accordance with YSI s written instructions ii abuse or misuse of the product iii failure to maintain the product in accord...

Page 3: ...nal Procedures 7 Measurement Modes or mg L 7 Calibration 8 Temperature 8 Dissolved Oxygen 8 Saving Viewing and Deleting Data 9 Downloading Data to a Computer DO200M Only 10 Recal Timer DO200M Only 10...

Page 4: ...tenance 13 Dissolved Oxygen Sensor 13 Temperature Sensor 14 Accessories Part Numbers 14 Technical Support 15 Specifications 15 606035REF B YSI DO200A DO200M User Manual English indd 4 12 7 2017 9 45 0...

Page 5: ...if properly installed and regularly maintained Erratic readings can result from damaged or fouled membranes or from large bubbles in the electrolyte reservoir If unstable readings or membrane damage o...

Page 6: ...40 DO200A meter manual 9V battery transport case polarographic probe with 10 meter cable 6 screw on cap membranes and electrolyte solution 601027 DO200M meter manual USB cable and 9V battery 601028 DO...

Page 7: ...play of BAT on the LCD indicates approximately one hour of battery life for unit operation within specifications Replace battery when BAT appears on the LCD To replace battery remove the two battery c...

Page 8: ...user prompts and mode indicators The unit prompts the user through calibration and measurement procedures The model DO200A DO200M field and lab probes use a polarographic electrode with convenient sc...

Page 9: ...isplay 8 Dissolved oxygen units or mg L 9 mbar Displays during calibration to prompt user for true barometric pressure 10 Save Full Recall Each All Delete Instrument s data storage indicators 11 Date...

Page 10: ...the DO200M these keys adjust recal time and select adjust date time format and information 3 DO200M INPUT EcoSense Mode CAL Recal Timer Date Time Calibration key Press to enter the calibration mode 4...

Page 11: ...9 2016 DO200M colors Black background PMS 7482 C power symbol PMS 130 C logo Reverse white all other imagery DO200M INPUT EcoSense Mode CAL Recal Timer Date Time key to turn the unit on or off The ins...

Page 12: ...y excess water to drain out of the bottle The wet sponge creates a 100 water saturated air environment for the probe which is ideal for calibration transport and storage of the DO probe For calibratio...

Page 13: ...ata record number for one second Full is displayed when trying to save data and memory is full To view saved data press Mode until Recall is displayed and then press Enter Use the Up or Down arrow key...

Page 14: ...o the Troubleshooting section 5 The instrument can be disconnected from the computer The original csv file should still be located on the DO200M instrument Recal Timer DO200M Only The Recal Timer feat...

Page 15: ...odel 200 BOD probe for the lab store the probe in a BOD bottle containing at least 1 inch of clean water Long term Storage For long term storage 30 days remove the battery from the instrument Moisten...

Page 16: ...ecimal DD MM YYYY The DO200M utilizes a decimal as the radix so a temperature of 31 1 is sometimes interpreted by Excel as 31 Jan when German is set as the PC language If a data file is opened in Exce...

Page 17: ...hen choose where you want the data to be placed on your opened spreadsheet M A I N T E N A N C E Dissolved Oxygen Sensor Clean the probe whenever it won t calibrate or the readings are unstable and ch...

Page 18: ...SI Temperature Sensor You must keep the temperature portion of the sensor free of build up Other than that the sensor requires no maintenance A soft bristle brush like a tooth brush can be used to scr...

Page 19: ...0 air saturation 0 1 air saturation 2 of the reading or 2 air saturation whichever is greater 0 00 to 20 0 mg L 0 01 mg L 2 of the reading or 0 2 mg L whichever is greater Pressure Compensation Input...

Page 20: ...Data DO200M only Waterproof USB for Downloading Data to PC DO200M only Conforms to the following Directives EMC 2014 30 EU RoHS 2011 65 EU WEEE 2012 19 EU Harmonized Standards EN61326 1 2013 IEC 61326...

Reviews: