YSI EcoSense DO200A User Manual Download Page 11

7

5

Mode

 key. Adjusts between 

measurement modes (%, 
mg/L), recalling saved data 
and deleting saved data. 

7

YSI EcoSense Handheld Overlay

12/19/2016

 

DO200M

colors:

Black background

PMS 7482 C power symbol

PMS 130 C logo

Reverse white all other imagery

DO

200

M

INPUT

EcoSense

®

Mode

CAL

Recal

Timer

Date

Time

Enter 

key. Pressing Enter saves the 

current measurement into memory, 
confirms mode selection (recall/
delete), confirms calibration steps, 
and confirms data deletion. On the 
DO200M, this key confirms recal 
timer entry and date/time selections.

8

YSI EcoSense Handheld Overlay

12/19/2016

 

DO200M

colors:

Black background

PMS 7482 C power symbol

PMS 130 C logo

Reverse white all other imagery

DO

200

M

INPUT

EcoSense

®

Mode

CAL

Recal

Timer

Date

Time

Date/Time 

key (DO200M only). 

A short press (i.e. key is not held) 
of the Date/Time key changes the 
display in the upper right corner to 
be either Date or Time. Pressing and 
holding for 3 seconds will allow date 
and time information to be updated. 
Pressing and holding for 6 seconds 
will allow for the date (e.g. MM/
DD/YYYY) and time format (12 or 
24 hour) settings to be changed. 

O P E R A T I O N A L   P R O C E D U R E S

Press the Power 

YSI EcoSense Handheld Overlay

12/19/2016

 

DO200M

colors:

Black background

PMS 7482 C power symbol

PMS 130 C logo

Reverse white all other imagery

DO

200

M

INPUT

EcoSense

®

Mode

CAL

Recal

Timer

Date

Time

 key to turn the unit on or off. The instrument will perform 

a self-diagnostic test, during which an “ovEr” message may appear on the 
display. This is normal. After the warm up is complete the message will 
disappear. After the self-diagnostic test completes, the temperature 
displays in the lower right of the display, and the unit is ready for operation. 
Immerse the probe halfway into the sample solution. If possible, do not 
allow probe to touch any solid object in the solution. Allow no air bubbles 
around the probe. When the unit is not in use, turn it off to save battery life. 
The instrument has a 30 minute auto shut off feature when not in use.

Note:

 During an oxygen measurement, the probe must be 

moved approximately 1/2 ft per second to overcome the inherent 
consumption of oxygen by the sensor. When using the 200-BOD 
lab probe, however, simply use the probe’s self-stirring feature.

Measurement Modes - % or mg/L

This unit provides three distinct measurements: Temperature, Dissolved 
Oxygen in % saturation, and Dissolved Oxygen in mg/L (ppm). The 
dissolved oxygen measurement unit is indicated on the right side of the 
display. Press Mode to switch between % and mg/L. 

606035REF-B-YSI-DO200A-DO200M-User-Manual-English.indd   7

12/7/2017   9:45:11 AM

Summary of Contents for EcoSense DO200A

Page 1: ...USER MANUAL English Fran ais Espa ol Portugu s EcoSense DO200A EcoSense DO200M Dissolved oxygen temperature instruments User Manual DOCUMENT 606035REF...

Page 2: ...amage or failure caused by i failure to install operate or use the product in accordance with YSI s written instructions ii abuse or misuse of the product iii failure to maintain the product in accord...

Page 3: ...nal Procedures 7 Measurement Modes or mg L 7 Calibration 8 Temperature 8 Dissolved Oxygen 8 Saving Viewing and Deleting Data 9 Downloading Data to a Computer DO200M Only 10 Recal Timer DO200M Only 10...

Page 4: ...tenance 13 Dissolved Oxygen Sensor 13 Temperature Sensor 14 Accessories Part Numbers 14 Technical Support 15 Specifications 15 606035REF B YSI DO200A DO200M User Manual English indd 4 12 7 2017 9 45 0...

Page 5: ...if properly installed and regularly maintained Erratic readings can result from damaged or fouled membranes or from large bubbles in the electrolyte reservoir If unstable readings or membrane damage o...

Page 6: ...40 DO200A meter manual 9V battery transport case polarographic probe with 10 meter cable 6 screw on cap membranes and electrolyte solution 601027 DO200M meter manual USB cable and 9V battery 601028 DO...

Page 7: ...play of BAT on the LCD indicates approximately one hour of battery life for unit operation within specifications Replace battery when BAT appears on the LCD To replace battery remove the two battery c...

Page 8: ...user prompts and mode indicators The unit prompts the user through calibration and measurement procedures The model DO200A DO200M field and lab probes use a polarographic electrode with convenient sc...

Page 9: ...isplay 8 Dissolved oxygen units or mg L 9 mbar Displays during calibration to prompt user for true barometric pressure 10 Save Full Recall Each All Delete Instrument s data storage indicators 11 Date...

Page 10: ...the DO200M these keys adjust recal time and select adjust date time format and information 3 DO200M INPUT EcoSense Mode CAL Recal Timer Date Time Calibration key Press to enter the calibration mode 4...

Page 11: ...9 2016 DO200M colors Black background PMS 7482 C power symbol PMS 130 C logo Reverse white all other imagery DO200M INPUT EcoSense Mode CAL Recal Timer Date Time key to turn the unit on or off The ins...

Page 12: ...y excess water to drain out of the bottle The wet sponge creates a 100 water saturated air environment for the probe which is ideal for calibration transport and storage of the DO probe For calibratio...

Page 13: ...ata record number for one second Full is displayed when trying to save data and memory is full To view saved data press Mode until Recall is displayed and then press Enter Use the Up or Down arrow key...

Page 14: ...o the Troubleshooting section 5 The instrument can be disconnected from the computer The original csv file should still be located on the DO200M instrument Recal Timer DO200M Only The Recal Timer feat...

Page 15: ...odel 200 BOD probe for the lab store the probe in a BOD bottle containing at least 1 inch of clean water Long term Storage For long term storage 30 days remove the battery from the instrument Moisten...

Page 16: ...ecimal DD MM YYYY The DO200M utilizes a decimal as the radix so a temperature of 31 1 is sometimes interpreted by Excel as 31 Jan when German is set as the PC language If a data file is opened in Exce...

Page 17: ...hen choose where you want the data to be placed on your opened spreadsheet M A I N T E N A N C E Dissolved Oxygen Sensor Clean the probe whenever it won t calibrate or the readings are unstable and ch...

Page 18: ...SI Temperature Sensor You must keep the temperature portion of the sensor free of build up Other than that the sensor requires no maintenance A soft bristle brush like a tooth brush can be used to scr...

Page 19: ...0 air saturation 0 1 air saturation 2 of the reading or 2 air saturation whichever is greater 0 00 to 20 0 mg L 0 01 mg L 2 of the reading or 0 2 mg L whichever is greater Pressure Compensation Input...

Page 20: ...Data DO200M only Waterproof USB for Downloading Data to PC DO200M only Conforms to the following Directives EMC 2014 30 EU RoHS 2011 65 EU WEEE 2012 19 EU Harmonized Standards EN61326 1 2013 IEC 61326...

Reviews: