Installation recommendations
ADA-WA 1; ADA-WA 4
6
ba76071e01
02/2012
Special installation recommendations:
VisoTurb 700 IQ and ViSolid 700 IQ
Note
Please also observe the general basic information in the
I
NSTALLATION
chapter of the sensor operating manual. There
you will find important notes on optimizing the installation.
45 ° pipe installation
Pipe diameter at the installation location of at least DN 80.
The pipe should be straight and not tapered for a length
of at least 50 cm (VisoTurb 700 IQ) or 25 cm
(ViSolid 700 IQ) in the direction of observation. Angled or
tapered pipes can cause interference effects in the case
of low turbidity.
The front part of the sensor is pitched towards the flow.
Exception:
If there are high quantities of foreign bodies with fibrous
or large surfaces, as for example hairs, strings or leaves,
it may be of advantage to incline the sensor in the
direction of the flow so that the sapphire disc does not
face the flow.
Min. 80 mm
ESS/WA...
WA 700/...
Marking aid in
this direction
Flow
direction
ADA-WA1
straight pipe
VisoTurb: min. 50 cm
ViSolid: min. 25 cm