ADA-WA 1; ADA-WA 4
Installation recommendations
3
ba76071e01
02/2012
Installation recommendations
General installation recommendations
Basic types of
installation
The optimum installation location must be individually deter-
mined for the type of sensor and the application.
Influential factors
The table on the following pages contains recommendations
and the special characteristics of the individual sensors.
Flow
direction
Minimum distance: 80 mm
A
45 °
45 °
B
C
Good
Sufficient flow of the
sensor element
A, B
Burdening of the measuring medium with
long fibers - avoidance of them getting
caught on the sensor (rake effect)
C
Burdening of the sample with small stones
and abrasive particles - avoidance of any
damage to the sensor element
C
Avoidance of malfunctions of the optical
measurement (turbidity/total suspended sol-
ids) by captured air bubbles and particles
A, B
Avoidance of malfunctions of the optical
measurement (turbidity/total suspended sol-
ids) by wall surfaces
A, C