Display Format
Menus → System → Configure Aux →Aux 1-2→Display Format
Configure the Display format (0, 0.0, or 0.00) based on the resolution to be displayed
on the Run Screen.
Example - when using a total dissolved gas pressure sensor configure, the display
format as 0.0 - figure 4.6, page 114.
In the below screen shot, the Run Screen is displaying aux name “TGP,” suffix
“mmHg,” with display format “0.0”. The display value represents an input of 1.875
volts with 0-5VDC analog aux configuration. The min value = “-400”. The max value
= “+400”. The control/alarm low value is “-50”. Alarms are enabled for this aux input.
figure 4.7
Notes:
-
Values are displayed based on the display format resolution. Example - if the
display format is 0.0 and the min value configured is 1.88, then 1.9 will be the
value displayed in the min value menu.
-
5200A aux sensor logs record 0.00 resolution regardless of the selected resolu-
tion in the aux analog display format configuration.
Directions to configure E-Aux analog - Configure E-Aux menu:
1. Wire E-Aux analog device - see IOEM Safety and Installation Guide>Chapter
3 Installation and Wiring>IOEM input configuration>Wire analog input.
2. Refer to and follow directions at IOEM Safety and Installation
Guide>Chapter 4>Quick Start - IOEM Configuration to confirm correct
network connectivity;
3. Follow directions in the Quick Start section to configure IOEM channel
mode to 0-1v, 0-5v or 4/20mA.
4. Enter E-Aux menu
5.
Menus → System → Configure E-Aux →E-Aux 1-8
6. Use
▲
and
▼
to scroll and highlight submenus to configure.
Configuring the 5200A
7. Press
to select.
8. Configure E-aux system name - page 108, instrument - page 108, channel
- page 109, input type - page 110, suffix - page 112, min/max - page 112,
scale min/scale max -page 113 and display format - page 114 menus.
9. Press
to save configuration.
10. Go to Sensor Setup Menus → Sensor Setup E-Aux 1-8 - page 161 to
enable and configure set points, control, alarm, relays, and hold off time.
11. Press
to save configuration.
E-Aux Temperature input type
Menus → System → Configure E-Aux →E-Aux 1-8→Input Type→Temperature
Directions to configure E-Aux temperature Configure E-Aux menu:
1. Wire temperature device - see IOEM Safety and Installation Guide>Chapter
3 Installation and Wiring>IOEM input configuration>Wire temperature
input.
2. Refer to and follow directions at IOEM Safety and Installation
Guide>Chapter 4>Quick Start - IOEM Configuration to confirm correct
network connectivity;
3. Follow directions in the Quick Start section to configure IOEM channel
mode to temperature.
4. Enter E-Aux menu
Menus → System → Configure E-Aux →E-Aux 1-8
5. Use
▲
and
▼
to scroll and highlight submenus to configure.
6. Press
to select.
7. Configure E-aux instrument - page 108, channel - page 109, input type
- page 110 and display format (Fo or Co).
8. Press
to save configuration.
9. Go to Sensor Setup Menus → Sensor Setup E-Aux 1-8 - page 161 to
enable and configure set points, control, alarm, relays, and hold off time.
10. Press
to save configuration.
Configuring the 5200A