4110(W)

General information

5

ba76137d04

10/2017

2.2

Opening and closing the IDS plug connection

This section only applies to IDS plug variants (4110W).

Opening the

 

plug

connection

If necessary, clean the plug connection.

Open the locking device (step 1).

Use your thumb and index finger to press the clips of the connector 
together, and pull the connector out of the plug (step 2).

Closing the plug

connection

Make sure that the plug connection is completely dry and clean.

Align the guidance of the connector with the groove in the plug and insert 
the connector in the unlocked plug until it catches (step 3).

Close the locking device (step 4).

2

4

3

1

Plug connection
closed and locked

Guidance

Groove

Summary of Contents for 4110-3

Page 1: ...OPERATING MANUAL ba76137d04 10 2017 4110 W 4110 3 pH ELECTRODES WITH GEL ELECTROLYTE BEDIENUNGSANLEITUNG ...

Page 2: ...2 ba76137d04 10 2017 4110 Contact YSI 1725 Brannum Lane Yellow Springs OH 45387 USA Tel 1 937 767 7241 800 765 4974 Email environmental ysi com Internet www ysi com Copyright 2017 Xylem Inc ...

Page 3: ...last 10 cali brations The data is recalled by the meter when the sensor is connected and is used for measurement and for measured value documentation Storing the calibration data in the sensor ensures that the correct slope and asymmetry are automatically used if the sensor is operated with different meters On the other hand different calibrated sensors can be used with one meter without the need ...

Page 4: ...t the electrode to the meter Calibrate the electrode according to the operating manual of the meter and observe the following rules while doing so Please keep the watering cap It is required for the electrode to be stored Always keep the watering cap clean 4110 via the sensor cable to a free IDS connec tor on the meter 4110W via a connecting cable accessory to a free IDS connector on the meter or ...

Page 5: ...tep 1 Use your thumb and index finger to press the clips of the connector together and pull the connector out of the plug step 2 Closing the plug connection Make sure that the plug connection is completely dry and clean Align the guidance of the connector with the groove in the plug and insert the connector in the unlocked plug until it catches step 3 Close the locking device step 4 2 4 3 1 Plug c...

Page 6: ...on and make sure the contact between the junction and test sample is thorough The junction is in the area of the bottom end of the shaft see arrow Caution Only the shaft part of the electrode may be immersed For measurements in aqueous solutions provide approximately the same stirring conditions for measuring as for calibrating Subsequent calibrations The frequency of subsequent calibrations depen...

Page 7: ...ulfides iodides proteins High temperatures High changes in pH and temperature The warranty does not cover failure caused by measuring conditions and mechanical damage 5 Maintenance and cleaning Cleaning Remove water soluble contamination by rinsing with deionized water Other types of contamination have to be removed as follows while the contact time with the detergents should be kept as short as p...

Page 8: ...ture sensor Integrated NTC 30 30 k at 25 C 77 F Connection cable Lengths 4110 1 5 3 m 4110W 1 5 3 m Diameter 4 3 mm Smallest allowed bend radius Fixed installation 20 mm Flexible use 60 mm Plug type Socket 4 pins Shaft dimensions material Shaft length 120 mm Shaft diameter 12 mm Shaft material PPE PS 4110 W IDS plug Synthetic materials Glass fiber reinforced Noryl TPU TPC ET POM PVC PEEK PBT O rin...

Page 9: ...le junction 4110 Gel Fiber Yes Plastic shaft Yes 4110 3 Gel Fiber Yes Plastic shaft Yes Maintenance equipment Description Model Order no Reference electrolyte solution 250 ml to fill the watering cap KCl 3 mol l Ag free KCl 250 109 705Y Connection cable 4110 W meter Description Model Order no IDS connection cable 1 5 m IDS CABLE 1 5 903 850Y IDS connection cable 3 m IDS CABLE 3 903 851Y Radio conn...

Page 10: ...Disposal 4110 W 10 ba76137d04 10 2017 ...

Page 11: ...ct for service include the Product Return form with cleaning certification The form must be completely filled out for an YSI Service Center to accept the instrument for service The Product Return form may be downloaded at www ysi com and clicking on the Support tab Telephone 800 897 4151 937 767 7241 Monday through Friday 8 00 AM to 5 00 PM ET Fax 937 767 1058 Email environmental ysi com Mail YSI ...

Page 12: ...Contact Information 4110 12 ba76137d04 10 2017 ...

Page 13: ......

Page 14: ...and return water to the environment and we help people use water efficiently in their homes buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications expertise backed by a legacy of innovation For more information on how Xylem can help you go to www xyleminc ...

Reviews: