Troubleshooting
Section 9
YSI, Incorporated
Model 30/30M
28
SYMPTOM
POSSIBLE CAUSE
ACTION
10.
Main Display reads “LErr”
A. In specific conductance mode,
temperature exceeds the values
computed using user defined
temperature coefficient and/or
reference temperature
B. During calibration, temperature
exceeds the values computed using
user defined temperature coefficient
and/or reference temperature
A. & B. Adjust user defined
temperature coefficient and/or
reference temperature (See section
6 Advanced Setup)
11.
Main Display reads “Err”
Secondary Display reads “ra”
A. System has failed its RAM test check
procedure
A. Turn instrument OFF and back ON.
B. Return the system for service.
12.
Main Display reads “Err”
Secondary Display reads “ro”
A. System has failed its ROM test check
procedure
A. Turn instrument OFF and back ON.
B. Return the system for service.
13.
Secondary Display reads “ovr”
A. Temperature is > 95
o
C
A. Read solution of lower temperature
B. Return system for service
C. Replace Probe/Cable assy.
14.
Secondary Display reads “udr”
A. Temperature is < -5
o
C
A. Read solution of higher temperature
B. Return system for service
C. Replace Probe/Cable assy.
15.
Secondary Display reads “rEr”
A. Temperature jumper is set to
o
F and
reading is >199.9
o
F but < 203
o
F
A. Set jumper to read
o
C.
B. Return system for service
The Model 30M has additional error codes as follows:
SYMPTOM
POSSIBLE CAUSE
ACTION
1.
Readings on main display don’t
change
A. Meter is in recall mode.
A. Press MODE key to return to Normal
Operation
2.
Main Display reads “nOnE”
A. During recall, no data is currently
stored in memory.
A. Store data before attempting to recall
Summary of Contents for 30
Page 2: ......
Page 4: ...ii ...
Page 6: ...Introduction Section 1 YSI Incorporated Model 30 30M 2 ...
Page 16: ...Saving Data Section 4 YSI Incorporated Model 30 30M 12 ...
Page 24: ...Advanced Setup Section 6 YSI Incorporated Model 30 30M 20 ...
Page 28: ...Principles of Operation Section 7 YSI Incorporated Model 30 30M 24 ...
Page 38: ...Warranty and Repair Section 10 YSI Incorporated Model 30 30M 34 ...
Page 40: ...Warranty and Repair Section 10 YSI Incorporated Model 30 30M 36 ...
Page 42: ...Accessories and Replacement Parts Appendix B YSI Incorporated Model 30 30M 38 ...
Page 48: ...Conversion Chart Appendix D YSI Incorporated Model 30 30M 44 ...