33
Hygiene and maintenance – Handpiece Nosecone
Maintenance
Visually inspect to ensure that all contamination has been removed.
Visually inspect power supply and cords for damage. Do NOT add lubricant to the handpiece.
Point of Use
Remove excess soil with a disposable cloth of paper wipe.
Discard wipe after use.
Containment and Transportation
Handle with care.
Manual Cleaning
After every use, clean first with a reprocessing agent listed on Page 31. Use additional wipes as necessary to ensure 2 minute contact
time.
> Note the manufacturer’s specifications for the use of the reprocessing agent.
Rinse and brush off with demineralized water (<38°C / < 100°F)
Remove any liquid residues using an absorbent cloth. Use a 2nd cloth to complete a final wipe.