32
Manual cleaning and reprocessing
Point of Use
Remove nosecone and process under guidelines outlined on Page 33.
Remove excess soil with disposable cloth or paper wipe. Discard wipe after use.
Containment and Transportation
Handle with care.
Cleaning
Using a pre-wet wipe or generously spraying on a clean cloth, thoroughly wet the surface of the device, and keep the surfaces
wet for 2 minutes. Use additional wipes to ensure 2 minute contact time. Ensure all visible soil is removed. After allowing device
to air dry, inspect for any residual soil. Use additional wipes as necessary to remove.
Reprocessing
Using a pre-wet wipe or generously spraying on a clean cloth, thoroughly wet the surface of the device, and keep the device wet
for 6 minutes. A second wipe should be used to contact every crack and crevice on the device. Use additional wipes to ensure
6 minute contact time. Allow device to air dry.
Drying
When cleaning, wipe surfaces dry with a clean cloth. Allow surfaces to air dry.
Ensure that no fluids enter the handpiece or foot control.
Do not place the handpiece or foot control into reprocessing solution or an ultrasonic bath!
Hygiene and maintenance – Handpiece, Foot Control, Charger/Adapter, and Cradle