![YOKOGAWA WU10 Instruction Manual Download Page 8](http://html1.mh-extra.com/html/yokogawa/wu10/wu10_instruction-manual_3444958008.webp)
IM 12B06W02-02E-E
8
Parameter pH diffeORP
pH differential is where the combined sensor has two dissimilar glass membranes
WU10-V-D, WE10 Color
EXA/FLXA terminal no. Description
Coax1 core
transparent
15
pH input
Coax1 shield
brown
16
pH guard
Coax2 core
transparent
13
Ref input
Coax2 shield
white
17
Ref guard
Wire
red
11
Pt1000-T1
Wire
blue
12
Pt1000-T2
Wire
yellow
14
Liquid Earth (LE)
Wire
green
Not connected
Drain wire
black
63 (4W analyzer only)
Shield
Parameter ORP (for pH differential)
WU10-V-D, WE10 Color
EXA/FLXA terminal no. Description
Coax1 core
transparent
Not connected
Coax1 shield
brown
Not connected
Coax2 core
transparent
13
Ref input
Coax2 shield
white
17
Ref guard
Wire
red
11
Pt1000-T1
Wire
blue
12
Pt1000-T2
Wire
yellow
15 & 14 (interconnected) Metal / Liquid Earth
(LE)
Wire
green
Not connected
Drain wire
black
63 (4W analyzer only)
Shield
Parameter ORP (pH compensated for pH differential)
WU10-V-D, WE10 Color
EXA/FLXA terminal no. Description
Coax1 core
transparent
13
Ref input
Coax1 shield
brown
17
Ref guard
Coax2 core
transparent
Not connected
Coax2 shield
white
Not connected
Wire
red
11
Pt1000-T1
Wire
blue
12
Pt1000-T2
Wire
yellow
15 & 14 (interconnected) Metal / Liquid Earth
(LE)
Wire
green
Not connected
Drain wire
black
63 (4W analyzer only)