![YOKOGAWA WU10 Instruction Manual Download Page 7](http://html1.mh-extra.com/html/yokogawa/wu10/wu10_instruction-manual_3444958007.webp)
IM 12B06W02-02E-E
7
Red (E)
Brown (B)
Transparent (A)
Yellow (G)
Blue (F)
White (D)
Transparent (C)
Green (H)
Black
Cable
A
B
C D E
F
brown
white
red
blue
brown
Connector
green
yellow
white
H G
3.2 Connections of VP dual coax cable WU10-V-D and WE10
Figure 3 Dual coax cable layout
Figure 4 VP- connector dual coax
Parameter: pH conveORP
pH conventional is where combined sensor with high ohmic pH membrane and low ohmic
internal reference.
WU10-V-D,
WE10
Color
EXA/FLXA terminal no. Description
Coax1 core
transparent
15
pH input
Coax1 shield
brown
13
Ref imput
Coax2 core
transparent
16
pH guard
Coax2 shield
white
14
Liquid Earth (LE)
Wire
red
11
Pt1000-T1
Wire
blue
12
Pt1000-T2
Wire
yellow
Not connected
Wire
green
Not connected
Drain wire
black
63 (4W analyzer only)
Shield
Parameter: ORP (for pH conventional)
WU10-V-D,
WE10
Color
EXA/FLXA terminal no. Description
Coax1 core
transparent
Not connected
Coax1 shield
brown
13
Ref input
Coax2 core
transparent
Not connected
Coax2 shield
white
15 & 14 (interconnected) Metal / Liquid Earth
(LE)
Wire
red
11
Pt1000-T1
Wire
blue
12
Pt1000-T2
Wire
yellow
Not connected
Wire
green
Not connected
Drain wire
black
63 (4W analyzer only)
Shield
Parameter: ORP (pH compensated, for pH conventional)
WU10-V-D,
WE10
Color
EXA/FLXA terminal no. Description
Coax1 core
transparent
13
Ref input
Coax1 shield
brown
Not connected
Coax2 core
transparent
17
Ref guard
Coax2 shield
white
15 & 14 (interconnected) Metal/Liquid Earth(LE
Wire
red
11
Pt1000-T1
Wire
blue
12
Pt1000-T2
Wire
yellow
Not connected
Wire
green
Not connected
Drain wire
black
63 (4W analyzer only)
Shield