Holding, Starting, and Stopping Integration
When the display is held, the integration operation continues, but the display and
transmission of the integration result are held. The relationship between the hold function
and the start and end operations is as follows:
• If you start integration while the display is held, the display and communication output
do not change. When you release the hold function (turned it off) or make a single
measurement by pressing SHIFT+HOLD (SINGLE), the integration result at that point
is displayed or transmitted.
ON
HOLD
OFF
Displayed value
Integration time
(The dotted line is the integrated value)
START
STOP RESET
• If you stop integration while the display is held, the displayed and transmitted values
do not change. When you release the hold function (turn it off) or make a single
measurement by pressing SHIFT+HOLD (SINGLE), the integration result at the point
when integration was stopped is displayed or transmitted.
Integration time
(The dotted line is the integrated value)
SINGLE
(Single measurement)
OFF
ON
HOLD
ON
ON
ON
Displayed value
START
STOP RESET
Sample Rate, and Valid Frequency Ranges for Integration
The sample rate is approximately 100 kHz. The voltage/current signal frequencies that
are valid for integration are as follows:
Integrated Item
Valid Frequency Range for Integration
Active power
DC to approximately 50 kHz
Current
When integrating Irms DC or the lower frequency limit determined by the
data update rate to approximately 50 kHz
When integrating Imn
DC or the lower frequency limit determined by the
data update rate to approximately 50 kHz
When integrating Idc
DC to approximately 50 kHz
When integrating Irmn DC or the lower frequency limit determined by the
data update rate to approximately 50 kHz
When integrating Iac
The lower frequency limit determined by the data
update rate to approximately 50 kHz
5.6 Integration