2-16
IM WT310E-01EN
2.6 Configuring Measurement Range Skipping
Procedure
Follow the procedure indicated by the thick lines in the following menu.
2.
6.
(Display D)
Select the measurement range mode.
(Display B)
1.
3.
7
.
Confirm
the setting.
Close menu.
4.
(Display C)
5
.
8.
Explanation
Turning Measurement Range Skipping On and Off
You can select which measurement ranges to skip (range configuration). When the auto range feature
is enabled, measurement ranges that are not used are skipped, and this instrument switches between
the measurement ranges that you have chosen to enable. Range skipping can reduce the loss of
measured data that can occur when ranges are switched one by one. The default setting is oFF.
• on: Range skipping is enabled.
• oFF: Range skipping is disabled.
Measurement Ranges That Are Skipped (Range configuration)
You can specify which measurement ranges to skip by sending communication commands through the
communication interface. You cannot specify which measurement ranges to skip from the display men
of this instrument. For details, see the Communication Interface User’s Manual, IM WT310E-17EN.
You can also specify which measurement ranges to skip using the WTViewerFreePlus software.
When using communication commands, select the enabled measurement ranges, not the ranges to
skip.
Peak Over Jump
In addition to measurement range skipping, you can set the measurement range to switch to when a
peak over-range occurs and the auto range feature is enabled. You can set this feature by sending
communication commands through the communication interface. You cannot set this feature from
the display men of this instrument. For details, see the Communication Interface User’s Manual, IM
WT310E-17EN. You can also specify which measurement ranges to skip using the WTViewerFreePlus
software.
• If set to OFF, the measurement range will increase in order through the enabled measurement
ranges when a peak over-range occurs.
• If measurement range skip is set to OFF, peak over jump will also not function.