5-7
IM WT310E-02EN
1
2
3
4
5
6
7
App
Index
External I/0 (Option)
Default Values for Integration: dFLt-i
Select this setting to output integrated values. The output settings are as follows:
Suffix Code
/DA4
/DA12
Product Name
WT310E
WT310EH
WT332E
WT333E
Output Channel
ch1
P1
P1
P1
ch2
WP1
-
P2
ch3
q1
P3
P3
ch4
fU
PΣ
PΣ
ch5
1
WP1
WP1
ch6
-
WP2
ch7
WP3
WP3
ch8
WPΣ
WPΣ
ch9
q1
q1
ch10
-
q2
ch11
q3
q3
ch12
qΣ
qΣ
1 These channels cannot be set.
2 The numbers are used to indicate input elements 1, 2, and 3.
Configuring an Original Output Format
You can specify output items (output functions and elements) for each output channel.
Output Functions (Area A in step 8 in the procedural explanation
for setting the D/A output format)
You can select from the following options.
u (voltage U), i (current I), P (active power P), VA (apparent power S),
VAr (reactive power Q), PF (power factor λ), dEG (phase angle Φ),
uFrq (voltage frequency fU), iFrq (current frequency fI),
uP (peak voltage value Upk), iP (peak current value Ipk),
Ph (sum of watt hours Wp), Ph+ (positive watt hour Wp+), Ph– (negative watt hour Wp–),
Ah (sum of ampere hours q), Ah+ (positive ampere hour* q+), Ah– (negative ampere hour* q-),
MATH (integration), ---- (0 V D/A output; no element is specified)
* For more information about positive and negative ampere hours, see page 5-4 of the User’s
Manual, IM WT310E-01EN.
Elements (Area B in step 10 in the procedural explanation for setting the D/A output
format)
Product Name
Element
WT310E or WT310EH 1
WT332E
1, 3, or 4
WT333E
1, 2, 3, or 4
Element number 4 represents Σ.
5.3 Producing D/A Output