6 - 25
IM 11G02M01-01E
6.7 Parameter setting
It allows you to carry out the parameter setting such as time, key lock, etc., as required. Items to be
set are as follows:
* For the maintenace mode, see Item 6.8.
1. To display the User mode, press the
MODE
key in the measurement mode.
2. Point the cursor to “Parameter Setting”
by pressing the
or
key. Press the
ENT
key.
3. In the “Parameter Setting” screen that
appears, point the cursor to any item you
want by pressing the
or
key.
Press the
ENT
key.
(
)
ENT
MODE
(
)
ENT
• Current Time
: Sets the current day of the week, hour and time.
Note: The clock backup time is 2 days. If power is turned on after
it is kept off for 2 days or louger, make the time setting again.
• Key Lock
: Sets with ON/OFF so that any key operation except the key
lock OFF cannot be performed.
• Remote Range
: Sets with ON/OFF whether the Range Switching is
made valid or invalid by external input.
• Output Hold
: Sets whether Calibration Output is held or not.
• Reset Average Output
: Resets the average value.
• Response Time
: Sets the response time of electrical system.
• Average Period
: Sets the moving average time.
• Maintenance mode
: Enters passwords to switch to the Maintenance mode.
Description of setting items