110
<Table 2>
After completing the calibration for ranges in Table 1, carry out calibration for “ Frequency
Characteristic”.
The calibration for is frequency characteristic required for AC voltage and AC current ranges
(marked with
).
Calibration is carried out at the designated frequency in table 2.
1) Set the calibrator to Input value as an input to the instrument.
2) Press the MEMORY key.
3) After 20 seconds, buzzer sounds and the instrument confirm calibration.
(Do not next key operation till buzzer sounds.)
Table 2. Input Signal for Calibration
Range
Input
Unit
AC 50mV*1
50.000
mV
AC 500mV
500.00
mV
AC 5V
5.0000
V
AC 50V
50.000
V
AC 500V
500.00
V
AC 1000V 600Hz
1000.0
V
AC 500μA
500.00
μA
AC 5000μA
5000.0
μA
5kHz
1kHz
*1: For model TY720.