99
To save a Data (LOGGING-mode)
In logging measurement mode, the time needs to be set. Note that changing the batteries resets
the time to 00:00. Set the time by referring to the Set-up function.
1) Press the MEMORY key. (The display shows the “ MEM ”symbol.)
2) Press the LOG (MIN/MAX) key.
(The display shows the logging interval (period).)
Set the value with the
(RANGE) key or
(REL
/
%) key.
The default setting is one second. (The default settings can be changed. Refer to the Set-up
function.)
The display shows “ FULL ” when the logging data is already saved.
When saving the new data, perform delete of data.
3) Press the LOG (MIN/MAX) key to start logging. (The “ MEM ”symbol is flashing.)
Every time the MIN/MAX key is pressed, the sub-display changes.
(Sub display : saved number → saved time (min : sec)→(hour : min)→saved number)
4) To cancel the function, hold down the MEMORY key for one second. When memory
capacity becomes full, the function is automatically canceled.
(“ MEM ” symbol disappears.)
Note
LOGGING-mode operation during HOLD-mode disables HOLD-mode.
To load a Data (SAVE-mode)
1) Press the MEMORY key. (The display shows the “ MEM ”symbol.)
2) Press the READ (LIGHT) key.
3) Press the SAVE (HOLD) key to select the number of saved data.
Select the number with the
(RANGE) key or
(REL /%) key.
4) To cancel the function, hold down the MEMORY key for one second.
(“ MEM ” symbol disappears.)
To load a Data (LOGGING-mode)
1) Press the MEMORY key. (The display shows the “ MEM ”symbol.)
2) Press the READ (LIGHT) key.
3) Press the LOG (MIN/MAX) key to select the number of saved data.
Select the number with the
(RANGE) key or
(REL
/
%) key.
Every time the MIN/MAX key is pressed, the sub-display changes.
(Sub display : saved number → saved time (min : sec)→(hour : min)→saved number)
4) To cancel the function, hold down the MEMORY key for one second.
(“ MEM ” symbol disappears.)
Summary of Contents for TY710
Page 9: ......