8-12
IM 704510-01E
8.5
Reading Measured Results Using Markers
For a functional description, see 2.6.
Procedure
Displaying Markers
1.
Press the MARKER key to display the Marker menu.
The menu item does not appear in inter-symbol interference mode since markers
cannot be used.
2.
Press the Marker soft key to select ON.
Selecting the Marker Control Item (only for dual measurement function)
3.
Press the Link soft key to display the Link menu.
4.
Press the Meas1, Meas2, or Link soft key to select the item to be scaled.
Selecting the Movement Speed of Markers
5.
Press the Movement soft key to select or
.
To decrease the marker speed, select ; to increase the speed, select
.
• Moving the X1 or X2 Marker Individually
6.
Press the X Marker soft key to select X1 or X2.
7.
Turn the rotary knob to select the marker.
The measured value and frequency at the marker position are shown below the
main window.
• Moving the X1 and X2 Markers Simultaneously
6.
Press the X1&X2 soft key.
7.
Turn the rotary knob to move the X1 and X2 markers simultaneously.
Explanation
Turning ON/OFF the Marker Display
Markers can be used to read out data. Readout values are displayed in the area below
the main window. The marker and readout value displays can be turned ON/OFF also
by using Style of the DISPLAY key. For dual measurement function, markers are
displayed on both Meas1 and Meas2 windows. Marker operation can be performed on
either window or both windows (Link)