3-18
IM 04P03B01-01E
3.13 Regarding the Message Format (/BT1 Option)
In the 5 message printouts, up to 35 characters can be printed out including the date/
time and measured values. Messages including measured values can be printed out
even when start printout or stop printout is disabled.
Procedure
The message format cannot be set by key operation. Set the format using the
communication command or a RXA10-03 or RXA10-04 Configuration software
(sold separately).
Explanation
Message Number
Select from 1 to 5.
Message Format
Choose a message format. Measured values, message strings, and date/time can be set
arbitrarily.
Example: Based on the message 1 string of the standard function, you can set a
combination with message format 1 and print it out as message 1 including
measured values.
10th and 11th character of the message
string (9 characters already used, so this
represents the next characters)
Standard function
Message 1 string:
Process-1
°
C
Message 1 printout result:
06/30 10:10 Process-1 134.8
°
C
Message format 1:
M/D H:M L09 01L02
Measured value on CH1 (no units)
Space (number of characters: 1)
Space
1-9th characters of the message string
Date/time
• Date/Time Format and Number of Characters Used
H:M
(Hour:Minute)
5 characters
H:M:S
(Hour:Minute:Second)
8 characters
M/D H:M
(Month/Day Hour:Minute)
11 characters
M/D H:M:S
(Month/Day Hour:Minute:Second)
14 characters
D/M H:M
(Day/Month Hour:Minute)
11 characters
D/M H:M:S
(Day/Month Hour:Minute:Second)
14 characters
D.M H:M
(Day.Month Hour:Minute)
11 characters
D.M H:M:S
(Day.Month Hour:Minute:Second)
14 characters
M.D H:M
(Month.Day Hour:Minute)
12 characters
M.D H:M:S
(Month.Day Hour:Minute:Second)
15 characters
Y/M/D H:M:S
(Year/Month/Day Hour:Minute:Second) 19 characters
M/D/Y H:M:S
(Month/Day/Year Hour:Minute:Second) 19 characters
D/M/Y H:M:S
(Day/Month/Year Hour:Minute:Second) 19 characters
D.M.Y H:M:S
(Day.Month.Year Hour:Minute:Second) 19 characters
M.D.Y H:M:S
(Month.Day.Year Hour:Minute:Second) 20 characters
• Measured Value Format and Number of Characters Used
The format for measured values is 01, 02, ...06. No units are added. 7 characters are