background image

23

IM 04L55B01-02EN

 

3.  Make that the source voltage is within the maximum rated 

voltage range of the provided power cord. Then, connect 

the other end of the cord to the outlet. Use a grounded 

three-prong outlet.

Precautions to Be Taken While Wiring the Power 

Supply (Power Supply Suffix Code: 2)

Make sure to follow the warnings below when wiring the 

power supply. Failure to do so may cause electric shock or 

damage to the instrument.

●  Wire the power cable to the power supply terminal, 

making sure that the polarity is correct.

●  Connect the power cables after checking that the 

power switch of the GM90PS is turned OFF.

●  Using other wires may cause abnormal heating or fire.

Wiring Procedure

1.  Turn off the GM90PS power supply, and then remove the 

transparent power supply terminal cover.

2.  Wire the power cable to the power supply terminal, making 

sure that the polarity is correct. 

Use ring-tongue crimp-on lugs (for M4 screws) with 

insulation sleeves. The recommended torque for tightening 

the screws is 1.4 to 1.5 N•m.

 

+

3.  Attach the transparent power supply terminal cover, and 

fasten it with screws.

Turning the Power On and Off

If the power switch of GM/GX60 systems under 

operation be turned on or off carelessly, it may result 

the system down or injury.

Check the following points before turning on the power 

switch.

●  The power cord or wires are connected properly.

●  The GM is connected to the correct power supply.

If the input wiring is connected in parallel with another 

instrument, do not turn on or off the GM or other 

instrument during operation. If you do, measured values 

may be affected.

Check the following points before turning off the power 

switch.

●  The GM10 is not accessing the external storage me

-

dium.

  You can turn the power on and off using the power switch 

on the front panel of the GM90PS.

  A self-test takes place for a few seconds, and then the GM 

will be running.

 

Power switch

Connecting GX90EX Expansion Modules

The GX90EX is used to configure a multi unit system.

•  For the main unit, link the GX90EX to the left end as seen 

from the front of the unit.

•  For a sub unit, link it next to the GM90PS.

Connect the GX90EXs of the main unit and sub units with 

Ethernet STP (shielded) cables. Only cascaded connection is 

supported.

Configuring the GX90EX Expansion Modules

Before setting the GX90EX dipswitches, turn off the unit.

 

Port

Setting switches (dipswitches)

System status LED

EXBus status LED

7 segment LED

Summary of Contents for SMARTDAC+ GM10

Page 1: ...ail 14 Mounting on a Wall 14 Wiring 15 Precautions to Be Taken While Wiring 16 Removing andAttaching a Terminal Cover 16 Removing andAttaching a Terminal Block 16 Wiring Procedure 17 Internal dimensio...

Page 2: ...out prior notice as a result of continuing improvements to the instrument s performance and functions Every effort has been made in the preparation of this manual to ensure the accuracy of its content...

Page 3: ...ration or shock or pull out the card Data can become corrupt or permanently lost Only use Yokogawa SD cards Operation cannot be guaranteed with other brands of card When inserting the SD card into the...

Page 4: ...iption GM90PS Power Supply Module for SMARTDAC GM Type 1 Always 1 Area N General Supply voltage 1 100 240V AC 2 12 28 V DC 1 Power supply connection D Power inlet with UL CSA cable F Power inlet with...

Page 5: ...01 1 1GB Shunt resistor for M3 screw terminal 415940 1 250 0 1 415941 1 100 0 1 415942 1 10 0 1 Shunt resistor for clamp terminal 438920 1 250 0 1 438921 1 100 0 1 438922 1 10 0 1 GM10 Style Number Re...

Page 6: ...system A unit consisting of a GM10 a GM90PS and a GX90EX Up to six I O modules can be connected to a unit Up to six sub units can be connected via the GX90EX Sub Unit multi unit system A unit consist...

Page 7: ...not be detected GM10 1 If you want to use the DI of a GX90XD or GX90WD only a single module installed in the GM main unit can be used Do not link modules in a way that violates the specifications Doi...

Page 8: ...or both do so before setting the date and time Download the following software applications SMARTDAC Standard Hardware Configurator SMARTDAC Standard IP Address Configurator See the list of electroni...

Page 9: ...e IM 04L55B01 01EN The multi unit system support only cascaded connections If connected in a ring none of the sub units will be identified Link the GX90EX to the left end on a main unit and next to th...

Page 10: ...e Expand reduce button 2 Click a No The corresponding GM10 is enabled 3 Click Network Config and set the items on the address setting screen For details on the settings see the IP Address Configurator...

Page 11: ...er 1 and click the button under Channel set Channel number display The Group number 1 Channel set dialog box appears 3 Set AI channel 001 to 0010 to On and click OK The selected channels appear under...

Page 12: ...unication Reconfiguration dialog box appears click OK The setup data of the connected GM will appear 3 Set the items 4 On the menu bar click the Setting tab and then Send Settings 5 When a Communicati...

Page 13: ...hese kinds of circumstances wait for at least an hour before using the GM to acclimate it to the surrounding environment Well Ventilated Location To prevent overheating install the GM in a well ventil...

Page 14: ...ront DIN rail DIN rail Rear of GM Click Check that all the latches are securely fastened to the DIN rail Removing from the DIN Rail 1 Lower the latch on the rear panel of the GM using a flat blade scr...

Page 15: ...ed For all other signal cables use basic insulated cables dielectric strength of 1390 VAC To prevent electric shock attach the terminal cover after wiring and make sure not to touch the ter minals App...

Page 16: ...n the thermocouple input take measures to stabilize the temperature at the input terminal Always use the terminal cover Do not use thick wires which may cause large heat dissipation we recommend a cro...

Page 17: ...ection port and tighten the screw terminal Wiring to a GX90XA Analog Input Module Universal Low withstand voltage relay Electromagnetic relay Current mA type Terminal Diagram CH1 301 201 101 101 102 1...

Page 18: ...A B B b A 5 DC current input with an external shunt resistor Be careful because the DI wiring is different between level and contact Terminal Arrangement M3 screw terminal CH No Term No Symbol Term N...

Page 19: ...DI8 COM COM DI9 DI16 COM DI9 DI16 DI1 DI8 COM Wiring direction Wiring direction 11 19 11 19 21 29 21 29 Terminal Arrangement Term No Symbol Term No Symbol 21 DI9 11 DI1 22 DI10 12 DI2 23 DI11 13 DI3...

Page 20: ...CH2 CH3 CH4 CH5 CH1 CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10 CH6 CH7 CH8 CH9 CH10 11 20 11 20 21 30 21 30 M3 screw terminal Clamp terminal Wiring direction Wiring direction Terminal Arrangement Term No S...

Page 21: ...on Internal circuit Connecting to the RS 422 485 Connector C3 option Two wire system Four wire system Electric potential of the shield Shield FG SDB SG SDA RDB RDA Electric potential of the shield Shi...

Page 22: ...rovide a power switch double pole type on the power supply line to separate the GM from the main power supply Use labels to indicate that this switch is for cutting off the power supply to the GM and...

Page 23: ...Power On and Off If the power switch of GM GX60 systems under operation be turned on or off carelessly it may result the system down or injury Check the following points before turning on the power s...

Page 24: ...L Red Lights during an error Setting Switches Dipswitches Use the dipswitches to set the unit number of the expansion module 10 Mbps fixed mode and operation mode master slave Dipswitch Settings Dipsw...

Reviews: